Operation
Quick Tips
Push the Displays button on the front panel to open the Select Displays dialog box.
To add a display, double-click the icon.
To remove displays, do one of the following:
Select a display icon in the Selected Displays list, and then click
Remove
.
Double-click a display icon in the Selected Displays list.
Click the Close button in the upper right hand corner of the selected display.
You can mix displays from multiple application folders.
Common Information Messages
The following table lists some of the common information messages that might appear during normal operation.
Table 3: Common information messages
Message
Explanation and recommended action
Acq BW too small for current setup
The display needs a wider acquisition bandwidth than the current data record
contains. This can be due to any of the following reasons:
A display other than the one you intended has been selected. The selected
display has requested a smaller acquisition bandwidth to achieve a better
accuracy or dynamic range for its particular measurement.
Select the display that contains the message. Click
Run
if the instrument
is not already acquiring data.
Acquisitions are not running and the measurement now requests a wider
bandwidth than the last acquisition.
Click
Run
to perform a new acquisition with a wider bandwidth.
The data is from a recalled
fi
le.
There is no way to increase the acquisition bandwidth for saved data. You
must adjust the measurement settings so that less bandwidth is required.
Need swept acq or larger Acq BW
When any display requires a swept acquisition, the other displays are unable to
process the swept data. This is because the swept data is customized for the
selected display.
Select the display that you are interested in and the acquisition settings will
automatically change to meet the requirements of the selected display.
RBW increased to 100 kHz
The current Spectrum Length or Analysis Length is not long enough to allow
a narrower RBW
fi
lter.
If the Length control is set to Manual, try increasing it or setting it to Auto so
that the Analysis Length will increase to the required value.