Functional Check and Calibration
DC Calibration Fixture
The DC calibration
fi
xture is shipped with
the probe and is used to automatically check
and adjust the probe gain and offset to the
connected oscilloscope channel.
1.
The
fi
xture is powered and automated
by a TekConnect control box adapter
that connects to the oscilloscope Aux In
connector.
2.
Output connectors on the
fi
xture board
accommodate the coaxial adapters and
solder tips that are used with the probe.
3.
The Probe Cal signal from the front panel
of the oscilloscope connects through a
cable to the BNC connector on the
fi
xture.
A BNC cable is included with the probe.
Fixture Notes.
The control box should be connected to the
fi
xture before it is connected to the oscilloscope. Although any input
channel can be used, using the Aux In channel allows channels 1–4 to be used for the calibrated probe(s).
WARNING.
To avoid damaging the
fi
xture, only apply power to the
fi
xture when it is on a
fl
at, nonconductive surface. The bottom
of the
fi
xture board contains exposed circuitry.
P7600 Series TriMode Probes Quick Start User Manual
17