Performance Veri
fi
cation
Equipment Setup
The following tests use two oscilloscopes; use this procedure to set up and warm the equipment to test the probe. Wear the
antistatic wriststrap when performing these procedures.
1.
Connect the 80A03 TekConnect probe interface to channels 3 and 4 of the TDS8000 oscilloscope. (See Figure 22.)
2.
Connect the 80E04 module to the 80A03 TekConnect probe interface.
3.
Connect the 80E08 module to channels 7 and 8 of the TDS8000 oscilloscope.
4.
Connect a 50
Ω
termination to the Aux Output connector on the probe, and connect the probe to one of the 80A03
channels.
5.
Turn on both oscilloscopes and allow 20 minutes for the equipment to warm up.
6.
Photocopy the test record to record the performance test results. (See Table 9 on page 47.)
Figure 22: Preliminary test setup
34
P7313SMA Technical Reference
Summary of Contents for P7313SMA
Page 1: ...P7313SMA 13 GHz Differential Probe Technical Reference P071196800 071 1968 00...
Page 2: ......
Page 3: ...P7313SMA 13 GHz Differential Probe Technical Reference www tektronix com 071 1968 00...
Page 6: ......
Page 8: ...Table of Contents ii P7313SMA Technical Reference...
Page 12: ...Preface vi P7313SMA Technical Reference...
Page 34: ...Operating Basics 22 P7313SMA Technical Reference...