![Tektronix P6243 Instructions Manual Download Page 20](http://html1.mh-extra.com/html/tektronix/p6243/p6243_instructions-manual_1078095020.webp)
Operating Basics
10
P6243 Instructions
Ground Lead Length
When you are probing a circuit, you should always use as short a
ground lead as possible between the probe head and circuit ground.
The input capacitance of the P6243 and the inductance of the ground
lead results in a resonant circuit that can ring. Using a shorter ground
lead will minimize ringing. See Figure 2.
Low-inductance
Ground
Three-inch
Ground
Six-inch
Ground
Figure 2: Waveform Distortion from Ground Lead Length
Figure 3 illustrates the resistance, capacitance, and inductance that
the probe and ground lead adds to a circuit.