Getting Started
P6243 Instructions
7
Figure 1: Probe Compensation Connections
1.
Connect the probe to the oscilloscope.
2.
Set the oscilloscope to display the probe channel.
3.
If you intend to compensate the probe, connect a ground lead and
an SMT KlipChip to the GND terminal.
4.
Using a standard tip, hold the probe to the SIGNAL terminal.
5.
Press
AUTOSET
(or adjust the oscilloscope) to display the
waveform specified for the compensation output.
6.
If you intend to compensate the probe, refer to the user manual
for your oscilloscope and perform the calibration routine now.
7.
Disconnect the probe tip from the SIGNAL terminal and connect
it to the GND terminal. The oscilloscope display should be at the
ground reference.