Performance Verification
26
P5210 High Voltage Differential Probe Service Manual
7.
Set the output of the generator to 500 VDC.
8.
Enable the output.
9.
Check that the trace on the oscilloscope shifts less than
1.67 divisions (167 mV) from center.
10.
Disable the generator output.
WARNING.
To reduce the risk of electric shock, ensure the generator
output is disabled before modifying/disconnecting test setup or
connections since exposed metal may at a hazardous potential. It is
recommended that the generator output amplitude be reduced to
minimum prior to disabling the output.
11.
Disconnect all test equipment.
This completes the performance verification procedure.