Operating requirements
Use these requirements to set up your environment and power source for proper
instrument operation.
Environment requirements
CAUTION.
Damage to the instrument can occur if this instrument is powered on at
temperatures outside the speci
fi
ed temperature range.
Table 1: OM2012 environmental requirements
Parameter
Description
Operating
+10 °C to +35 °C (+50 °F to +95 °F)
CAUTION.
Damage to the instrument can occur if
this instrument is powered on at temperatures outside
the speci
fi
ed temperature range.
Temperature
Nonoperating
–20 °C to +70 °C (-4.0 °F to +158 °F)
Relative
Humidity
Operating
15% to 80% (No condensation)
Operating
To 2,000 m (6,560 feet)
Maximum operating temperature decreases 1 °C each
300 m above 1.5 km.
Altitude
Nonoperating
To 15,000 m (49,212 feet)
Ventilation
Top, right, left: 0 inches (0 mm)
Front, rear: 2 inches (51 mm)
Bottom: 0.5 inches (13 mm) (bottom clearance provided by product feet)
Cable
clearance
Provide at least 2 inches (51 mm) front and rear so that connected signal
and power cables are not damaged by sharp bends.
OM2012 Installation and Safety Instructions
1