Trigger Setup
Selecting Triggers
Trigger Type
Trigger Conditions
Edge
Trigger on a rising edge, a falling edge, or both edges,
as de
fi
ned by the slope control. Coupling choices are
DC, LF Reject, HF Reject, and Noise Reject.
Edge triggers are the simplest and most commonly
used trigger type, with both analog and digital signals.
An edge trigger event occurs when the trigger source
passes through a speci
fi
ed voltage level in the speci
fi
ed
direction.
Sequence
(B Trigger) (not
available with an
edge slope of
Both
)
Combine an edge A Event (Main) trigger with the B
Event (Delayed) trigger to capture more complex signals.
(See page 65,
Time.
After the A Event occurs, the trigger system waits
the speci
fi
ed amount of time, and then looks for the B
Event before triggering and displaying the waveform.
Events.
After the A Event occurs, the trigger system
looks for a speci
fi
ed number of B Events before
triggering and displaying the waveform.
Pulse Width
Trigger on pulses that are less than, greater than, equal
to, or not equal to a speci
fi
ed time. Additionally, you can
trigger when a pulse width is within or outside a range of
two different speci
fi
ed times. You can trigger on positive
or negative pulses. Pulse width triggers are primarily
used on digital signals.
Timeout
Trigger when no pulse is detected within a speci
fi
ed
time. The signal stays above or below (or either above
or below) a set value for a set amount of time.
Runt
Trigger on a pulse amplitude that crosses one threshold
but fails to cross a second threshold before recrossing
the
fi
rst. You can detect positive or negative (or either)
runts, or only those wider than, less than, greater than,
equal to, or not equal to a speci
fi
ed width. Runt triggers
are primarily used on digital signals.
MDO3000 Series Oscilloscopes User Manual
93
Summary of Contents for MDO3012
Page 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Page 2: ......
Page 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Page 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Page 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...