Appendix A: Warranted Speci
fi
cations
Appendix A: Warranted Speci
fi
cations
The Analog Bandwidth when the instrument is DC-50 coupled and the Bandwidth Selection if “Full”.
The limits stated below are for ambient temperature of
≤
30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Bandwidth
10 mV/div to
1 V/div
5 mV/div to
9.98 mV/div
2 mV/div to
4.98 mV/div
1 mV/div to
1.99 mV/div
1 GHz
DC to 1.00 GHz
DC to 500 MHz
DC to 300 MHz
DC to 150 MHz
500 MHz
DC to 500 MHz
DC to 300 MHz
DC to 150 MHz
350 MHz
DC to 350 MHz
DC to 300 MHz
DC to 150 MHz
200 MHz
DC to 200 MHz
DC to 150 MHz
√
Analog
bandwidth, 50
Ω
100 MHz
DC to 100 MHz
√
DC Balance
0.2 div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.2 div with the input DC-75
Ω
coupled and 75
Ω
terminated
0.25 div at 2 mV/div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.25 div at 2 mV/div with the input DC-75
Ω
coupled and 75
Ω
terminated
0.5 div at 1 mV/div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.5 div at 1 mV/div with the input DC-75
Ω
coupled and 75
Ω
terminated
0.2 div with the input DC-1 M
Ω
coupled and 50
Ω
terminated
0.3 div at 1 mV/div with the input DC-1 M
Ω
coupled and 50
Ω
terminated
All above speci
fi
cations are increased by (0.01 divisions)/( °C) above 40 °C
√
DC Gain
accuracy
±2.5% for 1 mV/div, derated at 0.100%/°C above 30 °C
±2.0% for 2 mV/div, derated at 0.100%/°C above 30 °C
±1.5% for 5 mV/div and above, derated at 0.100%/°C above 30 °C
±3.0% variable gain, derated at 0.100%/°C above 30 °C
Offset, position and the constant offset term must be converted to volts by multiplying by the appropriate
volts/div term.
The basic accuracy speci
fi
cation applies directly to any sample and to the following measurements; High,
Low, Max, Min, Mean, Cycle Mean, RMS, and Cycle RMS. The delta volt accuracy speci
fi
cation applies
to subtractive calculations involving two of these measurements.
The delta volts (difference voltage) accuracy speci
fi
cation applies directly to the following measurements;
Positive Overshoot, Negative Overshoot, Pk-Pk, and Amplitude.
The limits are as follows:
Measurement Type
DC Accuracy in Volts
Average of > 16 waveforms
±((DC Gain Accuracy) X |reading - (offset - position)| + Offset
Ac 0.1 div)
√
DC Voltage
Measurement
Accuracy, Average
Acquisition mode
Delta Volts between any two averages of
≥
16 waveforms acquired with the same
scope setup and ambient conditions
±(DC Gain Accuracy X |reading| + 0.05 div)
MDO3000 Series Oscilloscopes User Manual
193
Summary of Contents for MDO3012
Page 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Page 2: ......
Page 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Page 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Page 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...