Search badges are created by tapping the Search button. Use the displayed configuration menu to set the search criteria.
Search badges have < (Previous) and > (Next) Navigation buttons that open the Zoom mode and center the waveform in the
display at the position of the previous or next search mark in the waveform record. Search badge Navigation buttons are only
usable when acquisitions are stopped.
You can drag or flick a Search badge to the right, off the display, to delete it. You can bring the badge back by swiping in from the
right.
Signal clipping and badges
Clipping causes inaccurate amplitude-related measurement results. Clipping also causes inaccurate amplitude values in saved
waveform files. If a math waveform is clipped, it will not affect amplitude measurements on that math waveform.
WARNING. Clipping is caused by excessive or dangerous voltages at the probe tip and/or a vertical scale setting that is not
adequate to display the entire vertical range of the waveform. Excessive voltage at the probe tip can injure the operator and
cause damage to the probe and/or instrument.
This instrument shows a warning triangle symbol and the word Clipping in a Channel badge when a vertical clipping condition
exists. Any measurement badges associated with that channel also indicate a clipping condition.
To close the clipping message, change the vertical scale to show the entire waveform, disconnect the probe tip from the
excessive voltage source, and check that you are probing the correct signal using the correct probe.
Getting acquainted with your instrument
24
MDO32, MDO34 Installation and Safety Manual