Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-40
2470-901-01 Rev. A /
May
2019
:OUTPut[1][:STATe]
This command enables or disables the source output.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
0 (OFF)
Usage
:OUTPut[1][:STATe] <state>
:OUTPut[1][:STATe]?
<state>
Turn source off:
0
or
OFF
Turn source on:
1
or
ON
Details
When the output is switched on, the instrument sources either voltage or current, as set by
:SOURce[1]:FUNCtion[:MODE]
.
Example
:OUTP ON
Switch the source output of the instrument to
on.
Also see
ROUTe subsystem
The ROUTe subsystem selects which set of input and output terminals to enable (front panel or rear
panel).
:ROUTe:TERMinals
This command describes which set of input and output terminals the instrument is using.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
FRON
Usage
:ROUTe:TERMinals <location>
:ROUTe:TERMinals?
<location>
Use the front-panel input and output terminals:
FRONt
Use the rear-panel input and output terminals:
REAR