FAQs
How should I clean the front-panel display?
If you need to clean the front-panel LCD touchscreen display,
use a soft dry cloth. If necessary, use a cloth dampened with an
ammonia-free glass cleaner. Do not spray cleaning
fl
uids onto the
display. Additionally, it is strongly recommended that you use only
fi
ngers on the touchscreen. Do not use sharp objects, such as a
pen or pencil, to touch the touchscreen.
My data looks odd or is wrong. What should I do?
Verify the connections from the instrument to the test
fi
xture. Also
check the connections from the DUT to the test
fi
xture socket.
I cannot unplug the mini-triaxial cable (4200A-MTRX) from
the SMU. What should I do?
The mini-triaxial connector is a locking connector. To remove it,
pull the knurled part of the connector back.
A confi guration error has been detected and I cannot
launch Clarius. What should I do?
This occurs when the physical con
fi
guration does not
match the con
fi
guration de
fi
ned in KCon or when there are
communication problems between instruments and the
4200A-SCS.
To verify system confi guration:
1. Run
KCon.
2. Select
Validate Confi guration
.
This can also occur if the preampli
fi
er, RPM, or CVIV is
removed or reconnected.
To update the system confi guration:
1. Run
KCon.
2. Select
Upgrade Preamp, RPM, and CVIV Confi guration
.
Note that preampli
fi
ers are SMU speci
fi
c. For example, a
preampli
fi
er that is con
fi
gured for SMU1 cannot be connected
to SMU2.