![Tektronix KEITHLEY 2450-EC User Manual Download Page 49](http://html1.mh-extra.com/html/tektronix/keithley-2450-ec/keithley-2450-ec_user-manual_1078009049.webp)
Models 2450-, 2460-, & 2461-EC Potentiostats And Galvanostats User's Manual Section 4: Potential Pulse & Square Wave
077110402/November 2016
4-5
To adjust the Base Potential:
1. From the Pulse Settings screen, select
Base Potential
.
2. Enter a value.
Figure 67: Entering a Base Potential
3. Select
OK
.
Next, you will adjust the Peak Potential. This setting allows you to adjust the maximum potential that
the instrument will attempt to force, given that the generated current is within the programmed current
limit.
To adjust the Peak Potential:
1. From the Pulse Settings screen, select
Peak Potential
.
2. Enter a value.
Figure 68: Entering the Peak Potential
3. Select