Characteristics 

The schematic of the 237-TRX-T is shown in the following figure. 

Figure 2: 237-TRX-T schematic 

 

 

Working voltage 

600 V

PEAK 

center conductor to inner shield 

1300 V

PEAK 

center conductor and inner shield to outer shell 

Maximum working current 

1 A

PEAK

 

Operating environment 

0 °C to 50 °C, up to 70% relative humidity at ≤35 °C   

Contact resistance 

< 0.5 Ω 

Insulation resistance 

10

15

 

Ω, center conductor to inner shield (500 V test voltage, 23 °C at < 40% 

relative humidity) 

The outer shell of the triaxial connector is for protection from voltage on the center and inner 

shield conductors. Make sure the outer shell is always connected to earth ground or a 

properly grounded chassis. 

 

Never touch or change the connections when power is applied to the adapter. Always turn off 

test system power and discharge all capacitors before connecting or disconnecting the 

adapter. 

To prevent voltages from being exposed or connections from shorting together, make sure 

both ends of the adapter are properly connected before applying voltage. 

 