Remote Operation
Series 2268 850 Watt DC Power Supplies Reference Manual
162
2268S-901-01 Rev. A / January 2015
SAVE AND RECALL
The save and recall of user settings can be done using commands as well as at the front panel.
Executing the save and recall commands will have the same outcome as following the
procedure outlined in “Saving User Setting Memory Locations” on page 59 and “Recalling
User Setting Memory Locations” on page 60.
Commands:
[:]SYSTem[<channel>]:SAVE {<user settings slot>}
[:]SYSTem[<channel>]:RECall {<user settings slot>}
Where:
<user settings slot> is any integer between 1 and 3. The value corresponds to the setting to be
operated on.
SET ANALOG PROGRAMMING LEVEL
Setting the analog programming level is used to define the range from 0 to the level that will
be input into the analog programming lines to control the output. The command is equivalent
to the procedure described in “Analog Programming Mode” on page 76.
Commands:
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel [:VOLTage][:ISOLated]
{?|<level>}
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel [:VOLTage] {?|<level>}
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel
:CURRent[:ISOLated] {?|<level>}
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel
:CURRent {?|<level>}
Where:
<level> is the maximum voltage or resistance that will be input to the analog inputs. The
acceptable range is from 2 to 10 V and 2 to 10kΩ.
SET REMOTE PROGRAMMING INTERFACE
The remote source can be select using the following SCPI commands. These commands are
equivalent to the procedure on “Voltage-Controlled Voltage APG Setup” on page 77 and
“Voltage-Controlled Current APG Setup” on page 79.
Commands:
[:]SYSTem[<channel>]:REMote:SOURce[:VOLTage]
{?|LOCal|AVOLtage|ARESistive}
[:]SYSTem[<channel>]:REMote:SOURce:CURRent
{?|LOCal|AVOLtage|ARESistive}
Where:
LOCal indicates the value is controlled by the set point.
AVOLtage is used to set the output to be controlled by an analog voltage input.
ARESistive is used to set the output to be controlled by a resistor circuit.
IAVoltage is used to set the output to be controlled by the isolated analog voltage input.
IAResistive is used to set the output to be controlled by the isolated analog resistor circuit.