Calibration
2-6
Performing 2 VDC Calibration
6. After completing 2VDC calibration, the instrument will
display the following:
CONNECT 20 VDC
7. Set the DC calibrator output to +20.000000V.
8. Press ENTER, and note that the instrument displays the
calibration voltage:
INPUT = 20.0000000
(At this point, you can use the cursor and range keys to set
the calibration voltage to a value from 9.5 to 20.5V if your
calibrator cannot source 20V.)
9. Press ENTER. The instrument will display the
following message to indicate it is performing 20V DC
calibration:
Performing 20 VDC Calibration
Figure 2-2
Calibrator connections for DC volts and ohms portion of comprehensive calibration
Input HI
Output HI
Input
LO
Output
LO
Model 2002
5700A Calibrator
Note
: Use shielded cables to minimize noise.
Enable or disable calibrator external
sense as indicated in procedure.
Sense HI
Sense LO
Sense HI
Sense LO
3. Set the calibrator output to +2.000000V, and turn exter-
nal sense off. Wait at least three minutes for thermal
equilibrium.
4. Press ENTER, and note that the Model 2002 displays
the presently selected calibration voltage:
INPUT = 2.00000000 V
(At this point, you can use the cursor and range keys to set
the calibration voltage to a value from 0.95 to 2.05V if your
calibrator cannot source 2V.)
NOTE
For best results, it is recommended that
you use the stated calibration values
throughout the procedure whenever
possible.
5. Press ENTER. The instrument will display the follow-
ing during calibration: