Command Descriptions
Returns
<number of trg lvl crossings>,(<TS for trg lvl crossing>,)...deadtime...<number of
trg lvl crossings>(<TS for trg lvl crossing>,)...deadtime...and so forth.
The format is set by the
command, and the data in parentheses is sent if
ON is active.
MEASure:ARRay:TIError? (Query Only)
This command automatically performs TIE measurements on clock signals from a
prede
fi
ned collection of system frequencies: 4, 8, 15.75, 64 kHz or 1.544, 2.048,
5, 10, 27, 34, 45, 52 MHz
TIE is de
fi
ned as positive and increasing if the measured frequency exceeds the
reference frequency.
Group
Measurement
Syntax
MEASure:ARRay:TIError?
(array size)[,[<exp
value>[,<resol>],][(@1|(@2)]]
MEASure:ARRay:TSTAmp? (Query Only)
Time stamps are taken of all positive and negative trigger level crossings of the
selected input channel. The commands
MEASure:ARRay:<MeasuringFunction>?
and
CONFigure:ARRay:<MeasuringFunction>
automatically invoke
ON
to get the time stamp data, but when
is used
instead, you should normally let it be preceded by the
ON
command explicitly. Otherwise the TS
values are omitted. See Returned
format below.
Measurements are performed in groups of four TS results, two positive and two
negative, with no deadtime between the values. Deadtime between groups is
affected by pacing and interpolator calibration, down to 4
μ
s.
Measurement results of 0 indicate negative trigger level crossings, whereas
positive values indicate the number of positive trigger level crossings since the
last reset.
2
TS is the time stamp value in seconds since a certain start event that is not available for external control.
So the TS values can only be used for relative time measurements.
Group
Measurement
Syntax
MEASure:ARRay:TSTAmp?
(<array size>)[,(@1)|(@2)]
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-75
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...