Command Descriptions
Returns
NORM or SWAP
FORMat:SMAX
Sets or queries the upper limit for FETCh:ARRay? MAX command in number
of samples. The command is intended to set an upper limit for use with any
controllers or application programs that cannot read large amounts of data.
Group
Format
Syntax
FORMat:SMAX <Numeric value>
FORMat:SMAX?
Arguments
Integer
N
, where 4
≤
N
≤
10000
Returns
<Numeric value>
FORMat:TINFormation
This command turns on/off the time stamping of measurements. Time stamping is
always done at the start of a measurement with full measurement resolution, and
is saved in the measurement buffer together with the measurement result.
The setting of this command will affect the output format of the MEASure, READ
and FETCh queries. See the
query.
For FETCh:SCALar?, READ:SCALar? and MEASure:SCALar? the readout
consists of two values instead of one. The
fi
rst value is the measured value and
the second value is the timestamp value.
In FORMat ASCII mode, both the measured value and the timestamp value are
given as
fl
oating-point numbers expressed in the basic units (Hz or s).
In FORMat REAL mode, the result is given as an eight-byte block containing the
fl
oating-point measured value, followed by an eight-byte block containing the
fl
oating point timestamp value.
In FORMat PACKed mode, the result is given as an eight-byte block containing
the
fl
oating-point measured value followed by an eight-byte block containing the
timestamp value expressed as a 64-bit integer (int64), the implicit unit being ps.
When doing readouts in array form, with FETCh:ARRay?, READ:ARRay?, or
MEASure:ARRay?, the response will consist of alternating measurement values.
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-57
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...