Command Descriptions
The main purpose of this function is to make continuous measurements of
relatively long period times without losing single periods due to result processing.
A typical example is the 1-pps timebase output from GPS receivers.
Group
Measurement
Syntax
MEASure:ARRay:PERiod:BTBack?
(<array size>)[,(@1)|(@2)]
Arguments
<ARRAY SIZE>
sets the number of samples.
(@1)|(@2)
is the measurement
channel:
(@1)
means input A
(@2)
means input B
MEASure:ARRay:STSTamp? (Query Only)
A time stamp (TS) is taken of the trigger level crossing on the selected
input channel. The commands
MEASure:ARRay:<MeasuringFunction>?
and
CONFigure:ARRay:<MeasuringFunction>
automatically invoke
ON
to get the time stamp data, but when
is used
instead, you should normally let it be preceded by the
ON
command explicitly. Otherwise the TS
values are omitted. See Returned
format below.
The deadtime to the next TS is due to pacing and interpolator calibration and
can go down to 4
μ
s. The X register/counter records the number of trigger level
crossings.
Depending on the state of the command
, one or two
values are output for each TS. If
OFF
, only the content of the X register/counter at
the timestamp is output. If
ON
, both the X register/counter and the TS value are
read and output as two values, separated by a comma in ASCII and REAL mode.
1
TS is the time stamp value in seconds since a certain start event that is not available for external control.
Therefore the TS values can only be used for relative time measurements.
Group
Measurement
Syntax
MEASure:ARRay:STSTamp?
(<array size>)[,(@1)|(@2)]
Arguments
Array size is the number of TS. One TS can contain 1 or 2 numeric values
depending on the state of the
command.
2-74
FCA3000, FCA3100, MCA3000 Series Programmer Manual
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...