Command Descriptions
Group
Format
Syntax
FORMat:TINFormation Boolean
FORMat:TINFormation?
FREQuency:BURSt:APERture
Sets the time length within a burst during which the burst frequency is measured.
Group
Sense
Syntax
FREQuency:BURSt:APERture {<Numeric value>|MIN|MAX}
FREQuency:BURSt:APERture?
Arguments
<NUMERIC VALUE>
is a number between 2E-8 (20 ns) and 2 s.
Returns
<Numeric value>
FREQuency:BURSt:PREScaler[:STATe]
The burst frequency limit is 300 MHz if the prescaler is ON and 160 MHz if
it is OFF.
Group
Sense
Syntax
FREQuency:BURSt:PREScaler[:STATe] <Boolean>
FREQuency:BURSt:PREScaler[:STATe]?
Arguments
<BOOLEAN> = (1 | ON | 0 | OFF)
Returns
1 | 0
FREQuency:BURSt:STARt:DELay
Sets the burst start delay (the time length between the burst start and the actual
start of the burst measuring time). This parameter controls the point of time when
a measurement sample is taken.
2-58
FCA3000, FCA3100, MCA3000 Series Programmer Manual
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...