Command Descriptions
ABORt (No Query Form)
The ABORt command terminates a measurement. The trigger subsystem state
is set to idle-state. The command does not invalidate already
fi
nished results
when breaking an array measurement. This means that you can fetch a partial
result after an abort.
Aborts all previous measurements if *WAI is not used.
Group
Measurement
Syntax
ABORt
ACQuisition:APERture
Sets the gate time for one measurement.
Group
Sense
Syntax
ACQuisition:APERture {<Decimal value > | MIN | MAX }
ACQuisition:APERture?
Arguments
<DECIMAL VALUE>
is 20 ns to 1000s.
MIN
sets 20 ns and
MAX
sets 1000 s.
Returns
<Decimal value >
200 ms after SYST:PRES
10 ms after *RST
ACQuisition:HOFF
Sets the Hold Off function On or Off.
Group
Sense
Syntax
ACQuisition:HOFF <boolean>
ACQuisition:HOFF?
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-29
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...