Command Descriptions
STATus:OPERation? (Query Only)
Reads out the contents of the operation event status register. Reading the
Operation Event Register clears the register. (See Figure 2-3 on page 2-103.)
Group
Status
Syntax
STATus:OPERation?
Returns
<Decimal data> = the sum (between 0 and 368) of all bits that are true. See the
following table.
Bit number
Weight
Condition
8
256
No measurement
6
64
Waiting for bus arming
5
32
Waiting for triggering and/or
external arming
4
16
Measurement
STATus:OPERation:CONDition? (Query Only)
Reads out the contents of the operation status condition register. This register
re
fl
ects the state of the measurement process. See table below.
Group
Status
Syntax
STATus:OPERation:CONDition?
Returns
<Decimal data> = the sum (between 0 and 3953) of all bits that are true. See
the following table.
Bit number
Weight
Condition
11
2048
Computing statistics
10
1024
In limit
9
512
Using internal reference
8
256
Meas. stopped / Computing
statistics (in compatibility
mode)
6
64
Waiting for bus arming
2-102
FCA3000, FCA3100, MCA3000 Series Programmer Manual
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...