Command Descriptions
Bit number
Weight
Condition
5
32
Waiting for triggering and /
or external arming
4
16
Measurement started
0
1
Calibrating
Figure 2-3: Device status continuously monitored
STATus:OPERation:ENABle
Sets the enable bits of the operation status enable register. This enable register
contains a mask value for the bits to be enabled in the operation status event
register. A bit that is set true in the enable register enables the corresponding bit in
the status register. (See Figure 2-3 on page 2-103.)
An enabled bit will set bit #7, OPR (Operation Status Bit), in the Status Byte
Register if the enabled event occurs. (See page 3-2,
Power-on will clear this register if power-on clearing is enabled via
Group
Status
Syntax
STATus:OPERation:ENABle <Decimal data>
STATus:OPERation:ENABle?
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-103
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...