Command Descriptions
READ?
takes a new measurement and fetches the result without changing the
programming of the instrument.
READ:ARRay? (Query Only)
The
READ:ARRay?
query is identical to
Aborts all previous measurement commands if
is not used.
NOTE.
The Statistics with array readouts cannot be combined and if the individual
values in a block measurement have to be stored, make sure the default command
is OFF.
The
READ:ARRay?
query differs from the
query by reading
out several results at once after making the number of measurements
previously set up by
CONFigure:ARRay:<MeasuringFunction>
or
MEASure:ARRay:<MeasuringFunction>?
.
Group
Read
Syntax
READ:ARRay?
{<array size for FETCh>|MAX}
Arguments
<ARRAY SIZE FOR FETCH>
sets the number of measurement results in the array.
The size must be equal to or less than the number of measurements in the output
buffer. The maximum limit is 10000 due to the physical size of the output buffer.
MAX
means that all the results in the output buffer are fetched.
Returns
<data>[,<data>]
The format of the returned data is determined by the format commands
.
Examples
ARM:COUN 10; READ:ARRAY?
:
5
con
fi
gures the instrument to make an array of
10 standard measurements. The instrument is triggered and data from the
fi
rst
fi
ve
measurements are read out with the
query.
*RMC (No Query Form)
This command removes an individual macro.
Group
Common
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-97
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...