Command Descriptions
INFinity
makes the arm loop continue inde
fi
nitely, or until other
device-dependent parameters set limits.
ARM:DELay
This command sets a delay between the pulse on the selected arming input and the
time when the instrument starts measuring.
Range: 20ns to 2s, with 10 ns resolution.
Group
Arming
Syntax
ARM:DELay <Numeric value> | MIN | MAX
ARM:DELay?
Arguments
<Numeric value>
is a number between 20*10
–9
and 2.
MIN sets 0 which switches the delay OFF.
MAX sets 2 s.
Returns
<Numeric value>
Examples
ARM:DELAY 0.1
ARM:LAYer2 (No Query Form)
This command overrides the waiting for bus arm, provided the source is set to
bus. When this command is issued, the instrument will immediately exit the
wait-for-bus-arm state.
The instrument generates an error if it receives this command when the trigger
subsystem is not in the wait-for-bus-arm state.
If the Arming source is set to Immediate, this command is ignored.
Group
Arming
Syntax
ARM:LAYer2
Examples
ARM:LAYER2
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-31
Summary of Contents for FCA3000 Series
Page 2: ......
Page 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 135: ...Status and Events ...
Page 136: ......
Page 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 157: ...Programming Examples ...
Page 158: ......
Page 179: ...Appendices ...
Page 180: ......
Page 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Page 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Page 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...