background image

Acquisition

Waveform record

The instrument builds the waveform record
through use of the following parameters:

Sample interval: The time between
sample points.

Record length: The number of samples
required to

fi

ll a waveform record.

Trigger point: The zero time reference in
a waveform record.

Horizontal position: When horizontal
delay is off, the horizontal position is
a percentage of the waveform record
between 0 and 99.9 percent. The trigger
point and the horizontal reference are at
the same time in the waveform record.
For example, if the horizontal position is
50 percent, then the trigger point is in the
middle of the waveform record. When
horizontal delay is on, the time from the
trigger point to the horizontal reference
is the horizontal delay.

Interpolation

Your instrument can interpolate between the samples it acquires when it does not have all of the actual samples it needs to

fi

ll

the waveform record. Linear interpolation computes record points between actual acquired samples by using a straight line

fi

t.

Sin(x)/x interpolation computes record points using a curve

fi

t between the actual values acquired. Sin(x)/x interpolation

is the default interpolation mode because it requires fewer actual sample points than linear interpolation to accurately
represent the waveform.

MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series User Manual

39

Summary of Contents for DPO5034B

Page 1: ...Series Mixed Signal Oscilloscopes DPO70000C DX Series Digital Phosphor Oscilloscopes DPO7000C Series Digital Phosphor Oscilloscopes MSO5000B and DPO5000B Series Oscilloscopes ZZZ User Manual P0712980...

Page 2: ......

Page 3: ...lloscopes DPO70000C DX Series Digital Phosphor Oscilloscopes DPO7000C Series Digital Phosphor Oscilloscopes MSO5000B and DPO5000B Series Oscilloscopes ZZZ User Manual xx This document supports firmwar...

Page 4: ...e privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc TekScope TekConnect Wave Inspector and TekLink are registered trademarks of Tektronix Inc FastFrame OpenChoice iCaptu...

Page 5: ...taxes and any other charges for products returned to any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and...

Page 6: ......

Page 7: ...moving the power 10 Securing the oscilloscope 11 Connecting to a network 11 Adding a second monitor 12 Operating system restore 15 Installing the MSO5000B and DPO5000B hard drive 16 Getting acquainted...

Page 8: ...ing a Main and b Delayed triggers 75 Triggering with B Event Scan 78 Triggering on a parallel bus 80 Triggering on a serial bus 83 Triggering using Visual Triggers Visual Triggering 84 Setting up Acti...

Page 9: ...ecalling waveforms 155 Saving digital waveforms 156 Saving instrument setups 157 Recalling instrument setups 158 Saving measurements 159 Saving user masks 160 Saving histogram data 161 Saving timestam...

Page 10: ...187 P6616 General Purpose Logic Probe instructions 187 Product description 187 Connecting the probe to the oscilloscope 188 Connecting the probe to your circuit 189 Functional check 189 Typical applic...

Page 11: ...ctors are exposed While using this product you may need to access other parts of a larger system Read the safety sections of the other component manuals for warnings and cautions related to operating...

Page 12: ...it off and disconnect the power cord Clearly mark the product to prevent its further operation Before use inspect voltage probes test leads and accessories for mechanical damage and replace when dama...

Page 13: ...rming any service procedures To avoid electric shock Do not touch exposed connections Do not service alone Do not perform internal service or adjustments of this product unless another person capable...

Page 14: ...ndicates an injury hazard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to property in...

Page 15: ...Western Road Bracknell RG12 1RF United Kingdom 1 This product is intended for use in nonresidential areas only Use in residential areas may cause electromagnetic interference 2 Emissions which exceed...

Page 16: ...cognized testing laboratory listing UL 61010 1 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements UL 61010 2 030 Safety Requirements fo...

Page 17: ...re sheltered locations where neither temperature nor humidity is controlled The area is protected from direct sunshine rain or direct wind Pollution degree 4 Pollution that generates persistent conduc...

Page 18: ...Compliance information Mains overvoltage category rating Overvoltage category II as defined in IEC 61010 1 xii MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manual...

Page 19: ...bout recycling options check the Support Service section of the Tektronix Web site www tektronix com Mercury notification This product uses an LCD backlight lamp that contains mercury Disposal may be...

Page 20: ...SO70000C DX DPO70000DX DPO70000C DPO7000C MSO5000B and DPO5000B Series instruments can help you verify debug and characterize electronic designs Key features include 33 GHz bandwidth and 50 GS s on 4...

Page 21: ...ou can limit the bandwidth down to 500 MHz to optimize the signal to noise ratio Enhanced bandwidth is extended to the probe tip for some high performance probes and tips Record lengths up to 500 000...

Page 22: ...control sample rate and record length separately from horizontal scale An intuitive graphical user interface UI with online help that is built in and available on screen Internal removable disk stora...

Page 23: ...adapter MSO DPO70000C DX models only TCA BNC Four TekConnect adapters MSO DPO70000C models only TCA 292MM Four TekConnect adapters MSO DPO73304DX MSO DPO72504DX and MSO DPO72304DX models only TCA 292D...

Page 24: ...Option A2 161 0344 00 161 0104 07 161 0210 00 Australia Option A3 161 0346 00 161 0104 14 161 0211 01 North America 240 V Option A4 161 0104 08 Switzerland Option A5 161 0347 00 161 0167 00 161 0212...

Page 25: ...inches 292 mm 4 Before operating the instrument verify the ambient temperature 5 C to 45 C 41 F to 113 F 5 C to 45 C 41 F to 113 F 5 Verify the operating humidity 8 to 80 relative humidity with a max...

Page 26: ...gs 1 5 V relative to the termination bias 30 mA maximum 5 V absolute maximum input 1 2 V FS settings 10 V at 25 C derated to 8 4 V at 45 C limited by attenuator Maximum nondestructive input voltage to...

Page 27: ...S with peaks 20 V DF 6 25 300 VRMS CAT II with peaks 425 V at the BNC For 100 mV div derate at 20 dB decade above 100 kHz to 30 VRMS at 1 MHz 10 dB decade above 1 MHz For 100 mV div derate at 20 dB de...

Page 28: ...hniques to prevent electrostatic discharge to the instrument 1 Discharge the static voltage from your body by wearing a grounded antistatic wrist strap while connecting and disconnecting cables and Te...

Page 29: ...C DX and DPO70000C DX Series 100 240 VRMS 10 50 60 Hz or 115 VRMS 10 400 Hz CAT II 1100 VA DPO7000C Series 100 240 VRMS 10 47 63 Hz or 115 VRMS 10 400 Hz 550 Watts maximum MSO5000B and DPO5000B Series...

Page 30: ...Install your instrument DPO7000C Series MSO DPO70000DX and MSO DPO70000C Series MSO5000B and DPO5000B Series 8 MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manual...

Page 31: ...tall your instrument Powering off the instrument MSO DPO70000DX MSO DPO70000C and DPO7000C Series MSO5000B and DPO5000B Series MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manual...

Page 32: ...stall your instrument Removing the power DPO7000C Series MSO DPO70000DX and MSO DPO70000C Series MSO5000B and DPO5000B Series 10 MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manu...

Page 33: ...twork Connect your instrument to a network for printing file sharing internet access and other functions Consult with your network administrator and use the standard Windows utilities to configure the...

Page 34: ...a dual monitor configuration MSO DPO70000DX and MSO DPO70000C Series 1 Turn power off 2 Connect second monitor If using a VGA monitor with an MSO DPO70000C use a DVI to VGA adapter 3 Connect keyboard...

Page 35: ...0C models 1 Turn off power 2 Connect second monitor 3 Connect keyboard 4 Connect mouse 5 Turn on instrument power 6 Turn on monitor power DPO7000C Series MSO DPO70000DX MSO DPO70000C DPO7000C and MSO...

Page 36: ...0B models 1 Turn off power 2 Connect second monitor 3 Connect keyboard 4 Connect mouse 5 Turn on instrument power 6 Turn on monitor power MSO5000B and DPO5000B models 14 MSO DPO70000DX MSO DPO70000C D...

Page 37: ...n Starting Acronis Loader press F5 for Acronis Startup Recovery Manager NOTE To successfully complete the system restore you must use the Windows version of the Acronis software Using a generic MAC ke...

Page 38: ...ly upside down could damage the instrument 1 Verify that the instrument is off 2 Insert the hard drive into the instrument with the drive assembly chassis facing up 3 Secure the drive assembly using t...

Page 39: ...ur instrument Getting acquainted with your instrument Front panel MSO DPO70000DX MSO DPO70000C and DPO7000C 1 DVD CD RW drive 2 Front panel controls MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO50...

Page 40: ...and DPO7054C 7 Probe compensation output 8 Probe calibration output 9 Channel 1 4 input 10 Auxiliary Trigger input 11 Fast Edge output 12 DC Probe Cal output 13 Logic probe input 14 Ground terminal th...

Page 41: ...Series 1 Front panel controls 2 Ground terminal 3 Probe compensation output 4 Auxiliary Trigger input 5 Channel 1 4 inputs 6 Logic probe input 7 USB ports MSO5000B and DPO5000B Series MSO DPO70000DX...

Page 42: ...ector 7 GPIB port to connect to controller 8 Line Out connector for speaker NOTE Some instruments may have additional audio connectors 9 Mic connector for microphone 10 DVI D Video port 11 COM 1 and C...

Page 43: ...tor for microphone 6 USB host ports 7 DVI 1 Video port 8 COM 1 and COM 2 serial ports 9 Printer connection 10 Oscilloscope Only XGA Out video port to connect a monitor 11 Removable hard disk drive 12...

Page 44: ...d 6 PS 2 connector for mouse 7 USB host ports 8 RJ 45 LAN connector to connect to network 9 USB Device port 10 Probe Power input use 119 7465 xx power supply 11 External reference input 12 Auxiliary o...

Page 45: ...trols Status Quick reference to vertical selections scale offset and parameters 7 Readouts Display cursor and measurement readouts in this area Measurements are selectable from the menu bar or toolbar...

Page 46: ...on icons to reposition a waveform 4 Click the icon to assign the multipurpose knobs to waveform vertical position and scale 5 Drag across the waveform area to create a box for zooming enabling disabli...

Page 47: ...quisitions 7 Use to set the trigger parameters Push Advanced to display additional trigger functions The Arm Ready and Trig D lights show the acquisition status 8 Turn to adjust waveform intensity 9 T...

Page 48: ...acquisition parameters of all waveforms 7 Turn the inner zoom knob to control the zoom factor Turn the outer pan knob to scroll the zoom window through the acquired waveform Push the play pause button...

Page 49: ...nal input 13 Push to turn the touch screen on and off 14 Push to display the math reference bus or vertical setup menus 15 Push to turn off a displayed menu 16 Push to access the digital setup menu MS...

Page 50: ...stem select Help Contents and Index 2 Use the Contents Index Search or Favorites tab to select the topic and then click Display To navigate within the help system Click a button in a help window to na...

Page 51: ...r a shortcut menu right click anywhere in the graticule or on an object The shortcut menu is context sensitive and varies with the area or object where you right clicked Some examples are shown in the...

Page 52: ...ternal diagnostics pass 1 Power on the instrument 2 Select Instrument Diagnostics 3 Click Run The test results appear in the diagnostics control window 4 Verify that all tests pass If diagnostic failu...

Page 53: ...st practice to run SPC when using the instrument to measure signals with higher sensitivity 10 mV div and lower settings regardless of temperature shift or time since it was last run Failure to do so...

Page 54: ...lified service personnel NOTE To always show the SPC status icon or display a warning when SPC has not been run in more than a month click the corresponding check box 5 If the SPC needed icon is red p...

Page 55: ...when making connections to the instrument and observe the maximum input voltage ratings for input connectors MSO DPO70000DX and MSO DPO70000C Series DPO7000C Series MSO5000B and DPO5000B Series 2 Sele...

Page 56: ...X MSO DPO70000C and DPO7000C Series 5 Adjust the horizontal position and scale using the front panel knobs The horizontal position determines the number of pretrigger and posttrigger samples MSO5000B...

Page 57: ...up 1 To quickly return to the factory default settings push DEFAULT SETUP MSO DPO70000DX MSO DPO70000C and DPO7000C Series MSO5000B and DPO5000B Series MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DP...

Page 58: ...robe and then select the input channel See page 33 Setting up analog signal input 2 Push the AUTOSET button to execute an Autoset MSO DPO70000DX MSO DPO70000C and DPO7000C Series MSO5000B and DPO5000B...

Page 59: ...compensation and deskew To optimize measurement accuracy see the instrument online help to perform the following procedures Compensate passive probes Compensate active probes Deskew input channels Ac...

Page 60: ...The instrument uses equivalent time sampling to extend its sample rate beyond its real time maximum sampling rate Equivalent time sampling is only used if Equivalent Time is selected and the time base...

Page 61: ...e trigger point is in the middle of the waveform record When horizontal delay is on the time from the trigger point to the horizontal reference is the horizontal delay Interpolation Your instrument ca...

Page 62: ...lower bandwidth waveform Envelope mode finds the highest and lowest record points over many acquisitions Envelope uses Peak Detect for each individual acquisition Average mode calculates the average v...

Page 63: ...NOB On or Off You will find this control in the Acquisitions tab of the Horizontal Acquisition control window Enhanced Effective Number of Bits enables interleave correction DSP that improves the effe...

Page 64: ...Click Mode and then select an acquisition mode 3 For Average or Envelope acquisition modes click the of Wfms control and then set the number of waveforms with the multipurpose knob For WfmDB mode clic...

Page 65: ...efault mode Select the horizontal mode that works best for your test setup To set the horizontal mode select Horiz Acq Horizontal Acquisition Setup to display the horizontal control window Choose one...

Page 66: ...ple Rate mode you can set the Sample Rate and Scale The default sample rate ensures bandwidth filter operation Record length is a dependent variable The maximum record length depends on your instrumen...

Page 67: ...ws that as you increase the scale the sample rate must decrease when either the maximum record length or the record length limit you set is reached Manual mode uses the maximum record length Automatic...

Page 68: ...000C Series MSO5000B and DPO5000B Series 2 Find glitches transients or other random events When you have identified an anomaly set the trigger system up to look for it See page 168 Capturing intermitt...

Page 69: ...sample rate The enhanced bandwidth provides a matched response across enabled channels so you can perform channel to channel comparison and differential measurements 1 Push AUTOSET to set the horizont...

Page 70: ...When different probing makes it impossible for the instrument to set all channels the same the instrument sets each channel to the closest bandwidth value possible The bandwidth indicator appears in...

Page 71: ...ows you to adjust the oscilloscope to mirror the conditions and behavior of the DUT and measure high speed signals in an environment similar to the one in which they operate Using the combination of t...

Page 72: ...k Detect or Hi Res acquisition mode 3 To stop acquisitions in Roll mode If you are not in Single Sequence push RUN STOP to stop Roll mode If you are in Single Sequence Roll mode acquisitions stop auto...

Page 73: ...be to the instrument 2 Connect the probe to the input signal sources CAUTION To prevent damage to the instrument always wear an antistatic wrist strap when making connections to the instrument and obs...

Page 74: ...ed 1 Select Digital Digital Setup 2 Select the D7 D0 or D15 D8 tab for the digital channels you want to set up 3 Click a digital channel button to toggle the display of the digital channel on or off 4...

Page 75: ...6 To change a digital channel label click the label and enter a new label using the keyboard Setting up a bus You can set up serial optional and parallel buses 1 Select Vertical Bus Setup or on some i...

Page 76: ...he bus click Bus Position and use the keyboard or multipurpose knob to enter a new bus position 7 To select the bus type click Bus Type Serial or Parallel You can set up each bus type See page 55 Set...

Page 77: ...up a bus 2 To select the input for the displayed components of the bus click the component Input and select from the displayed list 3 To set an input threshold double click Threshold and enter the th...

Page 78: ...for additional help setting up a bus 2 To add sources to the bus click the Add Sources Select button 3 Click the button for each channel you want in the bus The order of the parallel bus is determine...

Page 79: ...l Threshold and enter the threshold using the keypad or multipurpose knobs 9 When the desired thresholds have been set click the OK button 10 To set all channels to the same threshold double click Dig...

Page 80: ...up 2 Select the Display tab 3 To select the bus scroll to display the bus and then select the bus 4 Select the decoding from the bus decoding list See the online help for information on the Results Ta...

Page 81: ...ing packets physical bytes symbols and bit level 7 If using a symbol table check Use Symbol File Click Browse and browse to the symbol table file For some buses other settings are available Set the ot...

Page 82: ...trigger NOTE MagniVu centers itself around the trigger point If you turn MagniVu on while using a large record length and you are looking somewhere other than the trigger point then the digital signal...

Page 83: ...viewing the analog characteristics of a waveform iCapture is available on MSO70000C DX Series instruments and on DPO70000DX instruments with option MSOU installed 1 Select Vertical Vertical Setup 2 P...

Page 84: ...the list 6 For each threshold you want to set click the channel threshold Set the threshold using the keypad or multipurpose knobs 7 When the required thresholds have been set click the OK button Quic...

Page 85: ...he absolute trigger time for a specific frame and the relative time between triggers of two specified frames 1 Push AUTOSET to set the horizontal vertical and trigger controls or set the controls manu...

Page 86: ...frame that you want to view 6 To view multiple frames superimposed over each other select Overlay 7 Use the Time Stamps controls to select the source for and frame number of the reference frame The re...

Page 87: ...u use Temp or Spectral If you select an Average Summary Frame you can set Math functions Averaging to apply to All Frames or have it apply to Only the Summary Frame If the summary data is of primary i...

Page 88: ...lect the source of the FastFrame frames 2 Set the Search Range by entering the From Frame and the To Frame 3 Push Run Stop to stop acquisitions MSO DPO70000DX MSO DPO70000C and DPO7000C Series MSO5000...

Page 89: ...Finder searches for another different frame 6 If the anomaly you are looking for is in the displayed frame push Yes 7 If you are done searching push Stop 8 Push Run Stop to restart acquisitions MSO D...

Page 90: ...it is triggered If no trigger occurs the last waveform record acquired remains on the display If no last waveform exists no waveform is displayed Auto trigger mode enables the instrument to acquire a...

Page 91: ...ake the pretrigger period large enough to capture data before the glitch By analyzing what happens before the glitch you may uncover information that helps you find the source of the glitch Alternativ...

Page 92: ...ont panel or set up more advanced triggers in the Trigger Setup control window NOTE Some trigger type selections are not available on some instruments 1 Select Trig Edge Setup 2 Set the source slope c...

Page 93: ...directly from the Trig menu 4 Complete the trigger setup using the controls displayed for the trigger type The controls to set up the trigger vary depending on the trigger type instrument model and op...

Page 94: ...ive Serial Trigger on 64 bit NRZ serial pattern at data rates up to 1 25Gb s and 4 8b 10b symbols at data rates up to 6 25Gb s MSO70000C DX and DPO70000C DX only Requires option ST1G DPO7000C or ST6G...

Page 95: ...as a specified address of a bus that you define With some instruments and options trigger types include parallel SPI RS 232 USB and I2C triggers Not recommended with iCapture inputs if using more tha...

Page 96: ...e on a valid A event trigger has been recognized and the instrument is waiting for a delayed trigger When a delayed trigger is recognized the posttrigger part of the delayed waveform will fill If ARM...

Page 97: ...A Event occurs the trigger system looks for the B Event before triggering and displaying the waveform A trigger 1 Select Trig A Event Main Trigger Setup 2 Set the A trigger type and source in the A E...

Page 98: ...e Mode tab Trigger on B Event The A trigger arms the instrument Posttrigger acquisition starts on the nth B event B Trigger After Delay Time The A trigger arms the instrument Posttrigger acquisition s...

Page 99: ...the acquisition by an additional amount Correcting trigger position Trigger position correction corrects for differences in the data path and the trigger path to more accurately place the trigger on...

Page 100: ...e A Event for all acquisitions B Event Scan automatically advances the B Event value to capture different portions of the signal 1 Set the A trigger type and source in the A Event tab 2 Set the B trig...

Page 101: ...cted number of acquisitions occur Sequential Scan increments by 1 until the End Event is reached When the End Event is reached the B Event value is reset to the Start Event value and the process start...

Page 102: ...trigger on the DQS clock edges A B Sequence set to Trig on the nth Event B Scan Enabled with Start Event 1 End Event 8 and Mode Sequential The data eyes are formed by the DQ signal on Ch 2 Triggering...

Page 103: ...to trigger on NOTE Clocked buses will only appear in the drop down list if the clock source is set to Ch4 5 Click an Edit button and set the pattern and format that you want to trigger on MSO DPO70000...

Page 104: ...Pinpoint triggers 6 Select the data format 7 Set the pattern to trigger on using the keypad 8 Analyze your waveforms 82 MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manual...

Page 105: ...0000C DPO7000C MSO5000B and DPO5000B Series instruments Digital channels can be used in bus definitions on MSO instruments Set up a serial bus See page 53 Setting up a bus 1 Select Trig A Event Main T...

Page 106: ...gering Visual triggering allows you to create trigger conditions directly on the display screen Visual Triggers are available as an option on some models 1 Create a visual trigger area by left clickin...

Page 107: ...rea from the menu Use the visual trigger menus to edit the trigger area and set conditions for the visual trigger Refer to online help for detailed information about creating and editing visual trigge...

Page 108: ...s Select Save Send E mail or both Use the Setup button next to Save to display the Save on Event Setup display screen Use the Setup button next to Send E mail to display the Send E mail Setup display...

Page 109: ...the following procedure See page 88 Setting up E Mail on Event 1 Select Trig A Event Main Trigger Setup 2 Select the Mode tab 3 Under E mail on Trigger click On and then click Setup See page 88 Settin...

Page 110: ...mail address es Separate multiple entries with commas There is a limit of 252 characters in the e mail address box 3 Click Config and then enter the SMTP Server Address Contact your network administr...

Page 111: ...sage limit and e mail size When the maximum message limit is reached you must click Reset to send more e mails on event 7 To verify that you have set up the e mail addresses correctly send a test mail...

Page 112: ...time with the horizontal POSITION control or enter the delay time in the control window 4 Adjust the horizontal SCALE to acquire the detail that you need Quick tips Use MultiView Zoom and Horizontal D...

Page 113: ...the Display Style To set the display style select Display Display Style and then select one of the following styles Displays waveforms with lines drawn between record points Displays waveform record p...

Page 114: ...cquisition display settings Use for displaying points that may occur outside the normal acquisition envelope Variable persistence accumulates record points for a specified time interval Each record po...

Page 115: ...format to compare the voltage levels of the Ch 1 X and Ch 2 Y waveform records point by point as in XY format The displayed waveform intensity is modulated by the Ch 3 Z waveform record XYZ format is...

Page 116: ...fit between the actual samples acquired Linear interpolation computes record points between actual acquired samples by using a straight line fit Quick tips Sin X X interpolation is the default interpo...

Page 117: ...screen text 5 Click Controls to open the Text Properties control window for positioning of the text on the display 6 Click Clear to erase the entire text of the selected line Quick tips Click and dra...

Page 118: ...outs when cross hairs are not needed Use for making quick estimates of waveforms while leaving more room for automatic readouts and other data Use with automatic readouts and other screen text when di...

Page 119: ...of the graticule Long displays a horizontal line across the graticule Off turns off the trigger level marker Displaying the Date and Time 1 Select Display Objects 2 Toggle the display of the date and...

Page 120: ...shades The areas of lowest sample density appear in darker green shades It most closely resembles analog oscilloscope displays Spectral Grading displays areas of the waveform with the highest sample...

Page 121: ...lor for the math waveform as the lowest numbered channel waveform the math function is based on Quick tips The default color for math and reference waveforms are different for each waveform Using Mult...

Page 122: ...o display 3 To adjust the zoom graticule size select Zoom Graticule Size from either the Vertical or Horiz Acq menus 4 If MultiView Zoom is on but the readouts are not attached to the zoom controls pr...

Page 123: ...zoom 2 Select Zoom 1 On 3 Click and drag a box around another area of the waveform that you want to zoom and then select Zoom 2 On 4 To adjust the zoomed area horizontally click the horizontal marker...

Page 124: ...indow Turn each zoom display on and off from the Zoom Setup control window Push the MultiView Zoom button once or twice to toggle all zoom displays on and off To reposition the zoomed area horizontall...

Page 125: ...omed areas simultaneously click Lock and then select the Zoom1 4 check boxes that you want to scroll through Locking the zoomed areas locks in their relative horizontal position Changing the horizonta...

Page 126: ...Select a Zoom tab and then press Show Hide Waveforms 3 Select the zoomed area that contains the waveform you want to show or hide 4 Uncheck the channel math reference or digital waveform you want to...

Page 127: ...ctor is available on MSO5000B and DPO5000B Series instruments The Pan Zoom Control consists of 1 An outer pan knob 2 An inner zoom knob Zooming a waveform To use zoom 1 Rotate the inner knob on the Pa...

Page 128: ...Rotate the pan outer knob of the pan zoom controls to pan the waveform Turn the knob clockwise to pan forward Turn it counterclockwise to pan backward The further you turn the knob the faster the zoom...

Page 129: ...ions of the waveform You can mark areas of the waveform automatically if they meet some special criteria or you can manually mark each item of interest You can jump from mark to mark area of interest...

Page 130: ...r a search mark by turning a multipurpose knob Push the Next or Prev arrow button to jump to an existing mark 4 Push Set Clr or the front panel Set Clear button If no search mark is at the screen cent...

Page 131: ...another without adjusting any other controls 6 Delete a mark Push the Next or Prev arrow button to jump to the mark you want to clear To remove the center positioned mark push the Set Clr or the front...

Page 132: ...imilar to the trigger menu Serial bus searches are optional 3 Set up your search from the Configure tab To copy the trigger setup or a search setup press Settings Copy 4 From the Copy Settings window...

Page 133: ...earch 6 If not already on press All Searches to toggle search to on 7 On the screen triangles show the location of automatic marks and triangles with a white outline show the custom user defined locat...

Page 134: ...clear all marks from the table press All Marks Clear 11 To export the Mark Table to a file press All Marks Export 12 To convert a mark or all marks to user marks press Search Marks Save or Save All 13...

Page 135: ...ing Bring Zoom to Mark Zoom 2 or Zoom 3 displays the corresponding zoom view with the same zoom parameters as Zoom 1 Custom User marks are saved with the waveform when the waveform is saved and when t...

Page 136: ...binary or hex value I2C Search for Start Repeated Start Stop Missing Ack Address Data or Address and Data SPI Search for SS Active MOSI MISO or MOSI MISO CAN Search for Start of Frame Type of Frame D...

Page 137: ...up a visual trigger See page 84 Triggering using Visual Triggers Visual Triggering 3 Click Mark All Trigger Events in Record 4 Visual Trigger areas are displayed in blue 5 Visual Search areas are disp...

Page 138: ...his section contains concepts and procedures for analyzing waveforms Detailed information is available in the online help Taking automatic measurements 1 Select Measure Measurement Setup 2 Select the...

Page 139: ...an analog channel using iCapture See page 61 Viewing analog characteristics of digital waveforms In roll mode measurements are not available until after you stop the acquisition WARNING If there is v...

Page 140: ...between the maximum and minimum amplitude in the entire waveform or gated region Cycle RMS The true Root Mean Square voltage over the first cycle in the waveform or the first cycle in the gated regio...

Page 141: ...s and is measured over the entire waveform or gated region Histogram measurements Measurement Description Wfm Ct Displays the number of waveforms that have contributed to the histogram Hits in Box Dis...

Page 142: ...surement of the eye height in volts Eye Width The measurement of eye width in seconds Eye Top The top value used in the extinction ratio measurements Eye Base The base value used in the extinction rat...

Page 143: ...e levels Gating Use Gating to confine the measurement to a certain part of a waveform 1 Select Measure Gating Gating 2 Position the gates by doing one of the following Click Cursor to set the gated ar...

Page 144: ...e statistics select Off NOTE Placing the cursor over a measurement displays a higher resolution version of the measurement Snapshot To see a one time view of all valid measurements select Measure Snap...

Page 145: ...n from the Measurements setup control window From the drop down list select the measurement to annotate 2 To select the amount of measurement annotation detail select Measure Annotation Standard or De...

Page 146: ...lt High reference is 90 and Low reference is 10 Mid reference is primarily used for measurements between edges such as pulse widths The default level is 50 Mid2 reference is used on the second wavefor...

Page 147: ...s off MSO DPO70000DX MSO DPO70000C and DPO7000C Series MSO5000B and DPO5000B Series 2 Select the Cursor Source 3 Select a cursor type from one of the following H Bars measure amplitude typically in vo...

Page 148: ...ndem Use the Cursor Independent mode to move the cursors separately If you use the zoom graticule you can place a cursor directly on a specific waveform point to take precision measurements You can al...

Page 149: ...drag the pointer across the segment of the waveform that you want the histogram to cover For example make the box for a horizontal histogram wider than it is tall 2 Select Histogram Vertical or Histog...

Page 150: ...ow 5 You can also take automated measurements on histogram data See page 116 Taking automatic measurements Quick tips Use vertical histograms to measure signal noise and horizontal histograms to measu...

Page 151: ...ath waveforms you can derive the data view that your application requires Use the following procedure for predefined math equations 1 Select Math Math Setup 2 Choose one of the predefined math equatio...

Page 152: ...es and functions 4 When you have defined the expression to your satisfaction click Apply 5 To add your own filter click the Filter tab Click Load 6 Double click the folder for the filters you want to...

Page 153: ...Digital Properties to set the digital threshold of analog channels Measurements are taken on math waveforms in the same way as on channel waveforms Math waveforms derive their horizontal scale and pos...

Page 154: ...ath Math Setup 2 Choose one of the predefined spectral math expressions 3 Click Basic 4 Click Resolution BW or Frequency Span and use the keypad or the multipurpose knobs to adjust the spectral displa...

Page 155: ...ne 3 Click the type of spectral waveform that you want to create To redefine a waveform click Clear 4 Select the source waveform 5 Adjust the spectral waveform using the controls in the Spectral Setup...

Page 156: ...signal and increase the frequency resolution Different window functions produce different filter response shapes in the spectrum and result in different resolution bandwidths See the online help for...

Page 157: ...iver and margin testing The Error Detector has built in preset setups that cover the most common use cases These can be modified and saved as custom setup files for later recall Sample arbitrary wavef...

Page 158: ...The default location for Error Detector setup files is C Users Pub lic Tektronix TekScope ErrorDetector 4 Select the Setup tab 5 From the Source Setup tab select the Serial Bus Data Rate and signal So...

Page 159: ...G The Sync Pattern is a 10 20 30 or 40 bit unique pattern in the signal that the error detector uses to align the signal to the acquired comparison pattern The Advanced check box just allows you to se...

Page 160: ...require the learn operation Learn discovers the repeating pattern in the signal and stores it into system memory for comparison to the signal to determine errors in the signal Once a signal is in mem...

Page 161: ...illoscope memory and the Run operation starts the Error Detector testing for errors Note that the Learn operation only applies to some Bit error tests Remember Frame Symbol and Character error tests d...

Page 162: ...l This works because the Error Detector learn operation stores the signal test pattern into non channel specific memory 21 When the Error Detector is driven from the user interface bus triggers are us...

Page 163: ...the signal to pass the test it must fall outside the segments defined by the mask Generally standards committees such as ANSI define the masks To perform mask testing do the following 1 Select Mask So...

Page 164: ...ng a mask test 9 Click Autoset to automatically align the waveform with the mask based on the characteristics of the input signal 10 Toggle Autofit On to automatically reposition the waveform after ea...

Page 165: ...o run the mask test continuously 16 Select the notice you want when the test is complete 17 Click the Pass Fail Results tab to view the test results 18 Click Pass Fail Test On to start the mask test 1...

Page 166: ...selecting the Source Destination and Tolerances Use the multipurpose knobs to adjust the Tolerances Tolerances specify how much margin the signal is allowed before failing the limit test 3 Click Save...

Page 167: ...ck Highlight Hits On to show points that fall outside the template in a different color 10 Toggle Limit Test to On to start the test 11 Click Reset to clear all violations and reset the test Quick tip...

Page 168: ...res for creating and using MyScope control windows Detailed information is available in the online help Creating a new MyScope control window 1 Select MyScope New Control Window 2 Click to expand a ca...

Page 169: ...ol to preview it 4 Double click the control or click the to expand the control list If there is no then the control cannot be customized further MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B...

Page 170: ...lick and drag the control to your MyScope control window The control will snap to the nearest grid location when you release the mouse Change the placement of the control in your MyScope control windo...

Page 171: ...controls are deleted Select a control and then click Delete Just the selected control is deleted 11 Click Save and then enter a name for your MyScope control window or use the default name Quick tips...

Page 172: ...ntly used MyScope windows 2 Select the MyScope control window you want to use and then click Open To display the active MyScope control window do the following 1 Select MyScope Current or click MyScop...

Page 173: ...Quick tips Some controls function differently in a MyScope control window than they do in the standard control window For details see the online help You can copy MyScope control windows tcw files to...

Page 174: ...ements using the clipboard and printing on your instrument Detailed information is available in the online help Saving screen captures 1 Select File Save or Save As Screen Capture 2 Click Options to s...

Page 175: ...file type 5 Click Save Quick tip To quickly save multiple screen captures select Set Front Panel Print Button to Save and then click Save You can now save a screen capture by pushing the front panel P...

Page 176: ...waveform as a reference select Ref 1 4 To save as a wfm file select the location where you want to save the waveform 6 If you are saving as a wfm file type in a file name or use the default 7 Click S...

Page 177: ...veform to recall 5 Click Recall Clicking Recall turns on the reference waveform and activates the Reference Waveform control window 6 Use the controls to adjust the Reference waveform You can also acc...

Page 178: ...ve digital waveforms in a csv format file 1 To save digital waveforms select File Save or Save As Digitals 2 Click Digitals 3 You can only save digital waveforms as a csv file in a Windows directory T...

Page 179: ...the pop up keyboard to type a file name for a setup saved to the instrument memory 5 Click Save Quick tips If the touch screen is enabled use the pop up keypad to label the setups for easy identificat...

Page 180: ...of the ten locations in the instrument memory or from a Windows directory To retain your current deskew settings click Don t recall deskew values 4 Click Recall Quick tip You can recall any setup stor...

Page 181: ...ons to specify Displayed Measurements Measurement Snapshot or Measurement Format otherwise skip to step 3 3 Select the location to save the measurement 4 Type in a name for the measurement and then se...

Page 182: ...ks 1 Select File Save or Save As User Mask 2 Click User Mask 3 Select the location to save the mask 4 Type in a name for the mask and then select a file type 5 Click Save 160 MSO DPO70000DX MSO DPO700...

Page 183: ...gram Depending on previous selections you may need to select More Histogram Data to display the Histogram selection 3 Select the location to save the histogram 4 Type in a name for the histogram and t...

Page 184: ...need to select More Timestamp Table to display the Timestamp selection 3 Click Options to specify the Source or FastFrame Data Range otherwise skip to step 4 4 Select the location to save the timestam...

Page 185: ...o set the output content and format for images waveforms or measurements to be copied to the Microsoft clipboard 1 Select Edit Copy Setup 2 Click the Images Waveforms or Measurements tab and then sele...

Page 186: ...or measurements use the following procedure 1 Select the item that you want to copy 2 Select Edit Copy or press Ctrl C 3 Press Ctrl V to paste the item into a Windows application 164 MSO DPO70000DX M...

Page 187: ...lect File Print If necessary make changes to the page orientation in the Page Setup dialog box Your Print and Page Setup dialog boxes will depend on the printer you are using 2 Click Page Setup 3 Sele...

Page 188: ...s HT3 for HDMI analysis Use LT Waveform Limit Testing to compare an acquired waveform to tolerance boundaries that you set Use MOST Essentials for electrical compliance and debug testing for MOST50 an...

Page 189: ...g time measurements frequency and phase requires SVE Use USB to characterize USB 2 0 signals including mask testing and parametric testing Use USB3 to verify characterize debug and compliance test USB...

Page 190: ...t can be exceptionally tedious and time consuming to find infrequent anomalies especially given the low waveform capture rate on traditional digital storage oscilloscopes Digital Phosphor Oscilloscope...

Page 191: ...u look for the problem elsewhere go to step 4 4 Push FastAcq MSO DPO70000DX MSO DPO70000C and DPO7000C Series MSO5000B and DPO5000B Series 5 Find glitches transients or other random anomalies that are...

Page 192: ...hen set the width based on what you found in step 5 Check HF Filtering if you want the trigger system to treat a high frequency burst as a single pulse 10 Click E mail on Trigger On See page 88 Settin...

Page 193: ...work in real time To make your instrument the center of your design and documentation process use the following procedure NOTE On 64 bit systems you will need 64 bit compatible device drivers and appl...

Page 194: ...ent wherever you want to place the screen shot and then press Ctrl V Quick tips The instrument comes with a variety of OpenChoice software tools designed for maximum efficiency and connectivity to the...

Page 195: ...veforms NOTE Some trigger types are not available on some instruments For detailed bus triggering procedures use the following procedures See page 80 Triggering on a parallel bus and See page 83 Trigg...

Page 196: ...s you set up 7 Select Trigger On to select the desired trigger on feature 8 Depending on your Trigger On selection you may need to make additional selections 174 MSO DPO70000DX MSO DPO70000C DPO7000C...

Page 197: ...elds NOTE The video trigger type is available only on DPO7000C MSO5000B and DPO5000B Series instruments 1 Select Trig A Event Main Trigger Setup 2 Set the A trigger type and source in the A Event tab...

Page 198: ...e video lines in the field 1 Select Trig A Event Main Trigger Setup 2 Set the A trigger type and source in the A Event tab Select Format 525i NTSC 3 Select Trigger On All Lines 176 MSO DPO70000DX MSO...

Page 199: ...allows you to connect your logic analyzer to a Tektronix oscilloscope enabling communication between the two instruments The Add External Oscilloscope wizard which is available from the TLA applicatio...

Page 200: ...y switch using a dampened cleaning towel Do not spray or wet the switch directly CAUTION Improper cleaning agents methods or excessive force can damage the instrument Do not use chemical cleaning agen...

Page 201: ...in the Select Download Type pull down menu NOTE The Release Notes file available on the Web page may contain more recent information than the readme txt file contained within the downloadable executab...

Page 202: ...TPP1000 10X Passive Probes are compact passive probes with 10X attenuation that are designed for use with Tektronix MSO DPO4000B 5000B Series oscilloscopes The probes have no user or Tektronix service...

Page 203: ...signal on the oscilloscope when making this adjustment If the probe compensation fails verify that the signal and ground connections are secure at the Probe Comp connections on the oscilloscope Also...

Page 204: ...ix part number 206 0610 XX Pogo tip This spring loaded tip allows compliant testing of circuit boards Reorder Tektronix part number 206 0611 XX Ground lead with alligator clip Secure the lead to the p...

Page 205: ...06 0610 xx for rigid tip replacement or order part number 206 0611 xx for pogo pin replacement Specifications Table 1 Electrical and mechanical specifications Characteristic TPP0500 B TPP1000 Bandwidt...

Page 206: ...scription Temperature Operating Nonoperating 15 C to 65 C 5 F to 149 F 62 C to 85 C 80 F to 185 F Humidity Operating Nonoperating 5 to 95 relative humidity up to 30 C 5 to 45 relative humidity 30 C up...

Page 207: ...t or any products connected to it To avoid potential hazards use this product only as specified Using the probe or accessories in a manner not specified could result in a shock or fire hazard To avoid...

Page 208: ...sphere Keep product surfaces clean and dry Safety terms and symbols terms in this manual These terms may appear in this manual WARNING Warning statements identify conditions or practices that could re...

Page 209: ...digital buses and signals on your target system The probe contains 16 data channels split between two lead sets GROUP 1 and GROUP 2 The first lead in each set is identified by blue insulation and the...

Page 210: ...obe as shown in the illustration below 1 Insert the probe label side up into the connector on the oscilloscope 2 To remove the probe squeeze the buttons on the side and pull out the probe 188 MSO DPO7...

Page 211: ...n the Bus Setup menu On MSO DPO4000B Series instruments push the B1 to B4 buttons The setup screens allow you to set and view bus characteristics such as Clock type Bus type Serial or Parallel Bus wid...

Page 212: ...ground tip 1 set of 20 020 2711 XX 2 Probe tip 1 set of 10 131 5638 11 3 IC grabber 1 set of 20 020 2733 XX 4 Probe tip holder 2 ea 352 1115 XX 5 8 Ground lead 1 set of 2 020 2713 XX 6 3 Ground lead 1...

Page 213: ...skew 200 ps Probe length 1 0 m 3 28 ft Table 5 Environmental specifications Characteristic Description Temperature Operating 0 C to 50 C 32 F to 122 F Nonoperating 55 C to 75 C 67 F to 167 F Humidity...

Page 214: ...tions and components when power is present Avoid exposed circuitry Do not touch exposed connections and components when power is present Do not operate with suspected failures If you suspect there is...

Page 215: ...ktronix Inc Department or name if known 14200 SW Karl Braun Drive P O Box 500 Beaverton OR 97077 USA Email techsupport tektronix com Warranty information For warranty information go to www tektronix c...

Page 216: ...Appendix D Connecting the probe to the circuit 194 MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manual...

Page 217: ...tal channels analog characteristics 61 setting up 52 signal input 51 Digital waveforms saving 156 Display colors 99 objects 97 persistence 92 style 91 Display map 23 Display Remote 11 Documentation xv...

Page 218: ...22 Menus 29 MIPI CSI 2 173 MIPI DSI 1 173 Monochrome gray color palette 98 Monochrome green color palette 98 More measurements 119 Multiple zoom areas 101 MultiView zoom 99 MyScope editing 151 new con...

Page 219: ...alette 98 Termination voltge 49 Time measurements 118 Time Stamps 64 defined 63 Timeout trigger defined 72 Timestamp saving 162 Transition trigger defined 72 TRIG D status light 74 Trigger B Event sca...

Page 220: ...Index Z Zoom 99 105 graticule size 106 knob 105 Zoom graticule size 100 198 MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series User Manual...

Page 221: ......

Page 222: ......

Reviews: