Trigger Setup
For FlexRay, triggering occurs when the user-selected data input matches the data and quali
fi
er in the signal starting at the
byte offset. Set the number of bytes to match the number of bytes of interest. Use the data quali
fi
er to perform: =, !=, <, >, >=,
and <= operations. Triggering on identi
fi
er and data always matches the identi
fi
er and data selected by the user, with the
data starting at the
fi
rst byte. No rolling window is used.
RS-232 Bus Trigger Data Matching
You can trigger on a speci
fi
c data value for RS-232 bytes. If you de
fi
ned an end-of-packet character to use for RS-232 bus
decoding, you can use the same end-of-packet character as a data value for trigger data matching. To do so, choose the Tx End of
Packet or the Rx End of Packet character as the Trigger On selection.
Parallel Bus Trigger Data Matching
Optimum parallel bus trigger performance is achieved by using only analog channels or only digital channels (MSO4000 Series only).
MSO4000 and DPO4000 Series Oscilloscopes User Manual