DMM800 Series Performance Verification
16
Handheld Instruments Basic Service
b.
Using Wavetek 9100 with 9105 front porch:
�
Turn the calibrator output off.
�
Connect the multimeter test leads to the calibrator outputs.
�
Connect the calibrator common lead to the multimeter COM input.
�
Press the multimeter gold key followed by the
�
/% key.
�
Connect the remaining calibrator output lead to the multimeter
�
C V
�
�
input.
�
Turn the calibrator output on.
4.
Set the calibrator to each of the values in the Capacitance test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
5.
Turn the calibrator output off.
6.
Disconnect the calibrator from the multimeter.
Perform the following steps to verify the temperature measurement accuracy.
1.
Set the multimeter dial to
�
C /
�
F.
2.
Connect the ATK01 thermocouple adapter to the multimeter
�
C V
�
�
and
COM input connectors.
3.
Connect the Standard thermocouple (K type) of the calibrator to the ATK01
thermocouple adapter.
4.
Allow five minutes for the connector temperature to stabilize.
5.
Set the calibrator to each of the values in the Temperature test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
6.
Turn the calibrator output off.
7.
Disconnect the calibrator from the multimeter.
Perform the following steps to verify the DC volts peak measurement accuracy.
1.
Set the multimeter dial to V .
2.
Push the GOLD button and then the M/M/A button to select PEAK H.
3.
Connect the calibrator outputs to the multimeter
�
C V
�
�
and COM input
connectors.
Temperature Check
(DMM850 and DMM870)
Volts Peak Hold Check
(DMM870)