Section l-DM
SOlA
SPECIFICATION
Introduction
The
O
M
SOlA
Digital Multimeter measures dc vOltage
and current, ae voltage and current, d 8m, dBV, resistance,
and temperature. The ac functions are ac coupled only,
true rms responding. All the functions and ranges are front
panel push button selected, including the rear interface
connector input.
Readout in dBm or dBV is selected by an internal
jumper. The unit is shipped with the internal jumper in the
dBm position.
The readout is a
0.4"
high,
4 1/2
digit display using
seven segment LED. The decimal poi nt is automatically
positioned depending on the selected operating range of
the instrument. Polarity indication is automatic.
Accessories
Standard accessories include this instruction manual, a
set of test leads and the TEKTRONIX
P6601
temperature
probe with its instruction manual.
The
P6601
temperature probe and temperature
measurement capabilities are deleted for Option
2
In·
struments.
Performance Conditions
The electrical characteristics are valid only if the
OM
501A
has been calibrated at an ambient temperature
between
+210 C
and
+250
C and is operating at an ambient
temperature between
ooe
and
+50°C,
unless otherwise
noted.
Items listed in the Performance Requirements column
of the Electrical Characteristics are verified by completing
the Performance Check in the Calibration section at this
manual. These items are either explanatory notes or
performance characteristics for which
no
limits are
specified.
Table
1-1
ELECTRICAL CHARACTERISTICS (Front Panel)
Characteristics
Accuracy tor
200
mY,
2
V,
20
V
200
V and
1000
V ranges
+18°C
to
+2BoC
200
mV range
2
V to
200
V ranges
1000
V range
O°C
to
+ lBoC, +28°C
to
+50°C
200
mV to
200
V range
1000
V range
@
Performance Requirements
OC
VOLTMETER
±(0.05%
ot
reading
+ 0.015%
of full
scale)
±(0.05%
ot reading
+ 0.01%
of tull
scale)
±(O.OS%
ot reading
+ 0.02%
of full
scale)
±(0.1%
of reading
+ 0.025%
of full
scale)
±(0.1 %
of reading
+ 0.05%
of full
scale)
Supplemental Information
.
,
.
,
Summary of Contents for DM501A
Page 1: ......
Page 2: ......
Page 6: ...OM 501A OM SOiA Oigita Multlmeter xiv AUG 1980...
Page 13: ......
Page 18: ...Verso Filler Page z z...
Page 21: ...Operating Instructions OM SOlA FIg 2 2 Controls and connectors REV A AUG 1980 English 2 3...
Page 24: ...Verso Filler Page z z...
Page 25: ...OM 501A Japanese 2 3...
Page 26: ...Verso Filler Page z z...
Page 48: ...Verso Filler Page z z...