Reference
Reference
Operating system restore
The BERTScope analyzer contains an operating system restore
fi
le in a separate partition of the hard drive. Use this
fi
le
to restore the instrument operating system.
CAUTION.
Using the restore process reformats the hard drive and reinstalls the operating system. All saved data is lost. If
possible, save important
fi
les to external media before performing a system restore.
1.
Restart the instrument. During the boot-up process you will see the following message at the top of the screen:
Acronis Loader
Press F5 to start the Acronis Startup Recovery Manager.
2.
Repeatedly press the
F5
key until the Acronis True Image Tool opens. There is a 15-second delay after the message
appears until the instrument proceeds with normal instrument startup. If the instrument does not open the Acronis
application, power off the instrument, then power on the instrument and try again.
3.
Touch
Restore
.
4.
In the Con
fi
rmation dialog box, touch
Yes
to restore the instrument operating system, or
No
to exit the restore process.
The restore process takes approximately 30 minutes; the actual time depends on the instrument con
fi
guration.
5.
If you have not already done so, go to the Tektronix Web site to download the application software
(www.tektronix.com/software) and follow the instructions to install the software for your instrument.
20
Tektronix BERTScope Bit Error Rate Analyzers Quick Start User Manual