background image

Product disassembly procedures

Disassemble the instrument

1.

Disconnect the power cord from the instrument and set the instrument on a static-free, soft surface.

2.

Refer to the following

fi

gure and remove the six screws from the rear of the chassis, one screw from each corner and

two screws in the center of the rear panel.

Figure 4: Locations of screws on the rear panel

3.

Slide the instrument chassis out from the enclosure and set it aside.

14

Tektronix BERTScope BSA Series Declassi

fi

cation and Security Instructions

Summary of Contents for BERTScope BSA Series

Page 1: ...Tektronix BERTScope BSA Series Bit Error Rate Analyzer Declassification and Security ZZZ Instructions xx www tektronix com P077079100 077 0791 00 ...

Page 2: ...d and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc Contacting Tektronix Tektronix Inc 14150 SW Karl Braun Drive P O Box 500 Beaverton OR 97077 USA For product information sales service and technical support In North America call 1 800 833 ...

Page 3: ...and data export devices 7 Product disassembly procedures 13 Disassemble the instrument 14 Reassemble the instrument 15 Troubleshooting 16 Clear or sanitize a nonfunctional instrument 16 Recover from clearing or removing the instrument s memory 16 Change log 18 Tektronix BERTScope BSA Series Declassification and Security Instructions i ...

Page 4: ...ctric shock the grounding conductor must be connected to earth ground Before making connections to the input or output terminals of the product ensure that the product is properly grounded Observe all terminal ratings To avoid fire or shock hazard observe all ratings and markings on the product Consult the product manual for further ratings information before making connections to the product The ...

Page 5: ...s clean and dry Provide proper ventilation Refer to the manual s installation instructions for details on installing the product so it has proper ventilation Terms in this manual These terms may appear in this manual WARNING Warning statements identify conditions or practices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could result i...

Page 6: ...zard immediately accessible as you read the marking WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to property including the product The following symbol s may appear on the product iv Tektronix BERTScope BSA Series Declassification and Security Instructions ...

Page 7: ...d and resuscitation is present Disconnect power To avoid electric shock switch off the instrument power then disconnect the power cord from the mains power Use care when servicing with power on Dangerous voltages or currents may exist in this product Disconnect power remove battery if applicable and disconnect test leads before removing protective panels soldering or replacing components To avoid ...

Page 8: ...he procedures in this document are written to meet the requirements specified in NISPOM DoD 5220 22 M Chapter 8 ISFO Process Manual for Certification Accreditation of Classified Systems under NISPOM Products The following Tektronix products are covered by this document BSA85C BSA125C BSA175C BSA260C BSA286C Terms The following terms may be used in this document Clear This removes data on media mem...

Page 9: ...rce from the instrument Remove This is a physical means to clear the data by removing the memory device from the instrument Instructions are available in the product Service Manual Sanitize This eradicates the data from media memory so that the data cannot be recovered by other means or technology This is typically used when the device will be moved temporarily or permanently from a secured area t...

Page 10: ...Preface viii Tektronix BERTScope BSA Series Declassification and Security Instructions ...

Page 11: ...e type of information stored in the device Refers to instrument settings that can be changed by the user Both Describes the type of information stored in the device It means that both user data and user settings are stored in the device None Describes the type of information stored in the device It means that neither user data nor user settings are stored in the device Directly Describes how data ...

Page 12: ...nitize Fixed hard disk drive Holds operating system and application software Holds all user storable data patterns measurement results and instrument settings User data user settings Indirect Manufacturing operations user input Internal Yes Erase the hard disk drive with commercial erasure software Remove the hard disk drive LINK See page 10 Remove the hard disk drive When the hard disk drive is e...

Page 13: ...t functionality Flash 8 Mb Holds instrument manufacturing data User data user settings Indirect Firmware operations Generator stress module GSM board Yes N A does not contain user data or settings Clearing would disable instrument functionality N A does not contain user data or settings Sanitizing would disable instrument functionality Flash 16 Mb Holds instrument manufacturing data None Indirect ...

Page 14: ...settings Clearing would disable instrument functionality N A does not contain user data or settings Sanitizing would disable instrument functionality NVRAM Computer BIOS CMOS settings User settings User interface Written by processor system Computer board Yes See page 5 Clear or sanitize the CMOS settings Clearing would disable instrument functionality See page 5 Clear or sanitize the CMOS setting...

Page 15: ...instrument 1 Refer to the procedures to remove the chassis from the instrument enclosure See page 14 Disassemble the instrument 2 Locate the CMOS jumper J102 on the carrier board as shown in the following figure Figure 1 Location of J102 on the carrier board 3 Temporarily place a jumper across J102 to drain the residual voltage frm the CMOS memory Tektronix BERTScope BSA Series Declassification an...

Page 16: ...e page 15 Reassemble the instrument Clearing hard disk drives If your organization s security protocols allow the use of software to purge or sanitize hard drives you can use commercial software to erase free space on the hard drive before sending the instrument out for upgrades or repair Follow the instructions that come with the software to ensure that the erasure of sensitive data from the hard...

Page 17: ...ritten on the BERTScope or on a PC or USB flash drive can be removed or destroyed The USB ports can be disabled See page 7 Disable the RS 232 port and USB functions LAN Ethernet connector Control data Directly N A Rear panel The Ethernet port can be disabled See page 9 Disable the Ehternet and GPIB ports GPIB connector Control data Directly N A Rear panel The GPIB port can be disabled See page 9 D...

Page 18: ...the Integrated Peripherals menu go to USB Device Setting and then press Enter 12 Select USB IO Controller and then press Enter 13 Select Disabled and then press Enter 14 Select USB 2 0 Controller and then press Enter 15 Select Disabled and then press Enter 16 Press F10 and then select Yes to save and exit 8 Tektronix BERTScope BSA Series Declassification and Security Instructions ...

Page 19: ...er device for example Intel 82583V a Under the General tab go to Device usage at the bottom of the tab page and select Do not use this device disable b Click OK 6 In the Device Manager double click National Instruments GPIB Interfaces to expand the list 7 Double click PCI GPIB a Under the General tab go to Device usage at the bottom of the tab page and select Do not use this device disable b Click...

Page 20: ... the instrument enclosure See page 14 Disassemble the instrument 2 Locate the hard disk drive assemble on the top rear of the instrument 3 Loosen the screw securing the hard disk drive assembly using a P2 tip and then remove the assembly from the instrument Figure 2 Loosen the one screw to remove the hard disk drive assembly 4 Place the hard disk drive assembly on a clean static free working surfa...

Page 21: ...company s internal procedures for storing or disposing the hard disk drive 7 Reinstall the hard disk drive bracket without the hard disk drive in the computer card cage and tighten the one screw using a P2 tip 8 Refer to the procedures to reinstall the chassis in the instrument while being careful not to damage any circuit boards or cables See page 15 Reassemble the instrument Tektronix BERTScope ...

Page 22: ...Clear and sanitize procedures 12 Tektronix BERTScope BSA Series Declassification and Security Instructions ...

Page 23: ...avoid electric shock always power off the instrument and disconnect the power cord before cleaning or servicing the instrument CAUTION Many components within the chassis are susceptible to static discharge damage Service the chassis only in a static free environment Observe standard handling precautions for static sensitive devices while servicing the chassis Always wear a grounded wrist strap or ...

Page 24: ...ce 2 Refer to the following figure and remove the six screws from the rear of the chassis one screw from each corner and two screws in the center of the rear panel Figure 4 Locations of screws on the rear panel 3 Slide the instrument chassis out from the enclosure and set it aside 14 Tektronix BERTScope BSA Series Declassification and Security Instructions ...

Page 25: ...side the shell Verify that the EMI gasketing is seated on the front edges of the shell CAUTION To avoid damaging the power cord connector center the power connector in the hole as you tighten the screws 2 Attach the chassis to the cover shell using the six screws that you removed earlier Do not tighten the screws until all of them have started a Install one of the center screws while moving the ch...

Page 26: ...om clearing or removing the instrument s memory After the hard disk drive has been sanitized the operating system directory structure computer drivers and the BERTScope software must be reinstalled Replace the hard disk drive with a new preprogramed hard disk drive from Tektronix Customer Service Complete the following steps to reset the single board computer settings to the default values 1 Power...

Page 27: ...page 7 Disable the RS 232 port and USB functions 8 Select Save Changes and Exit and then press Enter twice The instrument automatically restarts 9 To disable the Ethernet and GPIB ports refer to the procedures earlier in this document Tektronix BERTScope BSA Series Declassification and Security Instructions 17 ...

Page 28: ...Change log Change log Document part number Revision date Change description 077 0791 00 January 10 2013 Initial release 18 Tektronix BERTScope BSA Series Declassification and Security Instructions ...

Reviews: