Troubleshooting
AWG7000 Series Service Manual
4-45
Table 4-6: Diagnostics error
Error code
Error log message
Possible defect
on board
Note
120000000
HwDiagAwgRegister AwgController address_0x21 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000001
HwDiagAwgRegister AwgController address_0x23 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000002
HwDiagAwgRegister AwgController address_0x30 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000003
HwDiagAwgRegister AwgController address_0x40 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000004
HwDiagAwgRegister AwgController address_0x41 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000005
HwDiagAwgRegister AwgController address_0x42 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000006
HwDiagAwgRegister AwgController address_0x50 register
diag. Fail expected data=xxxx actual data=xxxx
AWG10G
AWG10G: U1000
120000007
HwDiagAwgRegister Ch1 MemoryController-
N
Communication Fail expected data=xxxx actual data=xxxx
-N:PLD no
AWG10G: U1000
N=0 -> U400PA
N=1 -> U400QA
N=2 -> U400RA
N=3 -> U400SA
N=4 -> U400TA
120000008
HwDiagAwgRegister Ch2 MemoryController-
N
Communication Fail expected data=xxxx actual data=xxxx
-N:PLD no
AWG10G: U1000
N=0 -> U400PB
N=1 -> U400QB
N=2 -> U400RB
N=3 -> U400SB
N=4 -> U400TB
120000009
HwDiagAwgRegister Ch1 TEK0015-
TNO
address_
REG
register diag. Fail expected data=xxxx actual data=xxxx
TNO: TEK0015 number
REG: Register number
AWG10G: U1000
TNO=0 -> U300PA
TNO=1 -> U300QA
TNO=2 -> U300RA
TNO=3 -> U300SA
TNO=4 -> U300TA
120000010
HwDiagAwgRegister Ch2TEK0015-
TNO
address_
REG
register diag. Fail expected data=xxxx actual data=xxxx
TNO: TEK0015 number
REG: Register number
AWG10G: U1000
TNO=0 -> U300PB
TNO=1 -> U300QB
TNO=2 -> U300RB
TNO=3 -> U300SB
TNO=4 -> U300TB
120000011
HwDiagAwgRegister Ch1 HFD205 address_
REG
register
diag. Fail expected data=xxxx actual data=xxxx
REG: Register number
AWG10G
AWG10G: U1000, U100A
120000012
HwDiagAwgRegister Ch2 HFD205 address_
REG
register
diag. Fail expected data=xxxx actual data=xxxx
REG: Register number
AWG10G
AWG10G: U1000, U100B
121000000
HwDiagWaveformMemory channel=1, MemoryController-
P
Data bus Error
pinName
expected data=xxxxxxxx actual
data=xxxxxxxx
P: PLD number
pinName:RAM pin name
AWG10G
P=0 -> U400PA,U500PA, U510PA
P=1 -> U400QA,U500QA, U510QA
P=2 -> U400RA,U500RA, U510RA
P=3 -> U400SA,U500SA, U510SA
P=4 -> U400TA,U500TA, U510TA
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......