Troubleshooting
4-44
AWG7000 Series Service Manual
Instrument Diagnostics
The primary diagnostics for the instrument are accessible through the AWG
application software. The Instrument diagnostics check the basic functionality of
analog and marker output at every power on. The Instrument Diagnostics run after
Windows has booted up and run its own set of system checks. If any failures occur
at power on, a pop-up message indicates that a failure has occurred and provides
an error code. Table 4-6 lists diagnostics error codes, Table 4-7 lists calibration
error codes, and Table 4-8 lists hardware error codes.
The power-on tests ensure that the hardware is installed and can be accessed by the
software. The tests provide limited diagnostic information, and provide no
performance information.
To run the instrument Diagnostics manually, do these steps:
1.
Turn off all other applications.
2.
From the menu bar, select
System
and then select
Diagnostics...
; the
Diagnostics dialog box appears (see Figure 4-22).
Figure 4-22: Diagnostics dialog box
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......