Adjustment Procedures
1.
Install the test hookup and preset the instrument controls:
a.
Hook up the sampling oscilloscope:
Attach the SMA terminations to the Channel 1 Analog (–) Output,
Channel 2 Analog (–) Output, and Marker 1 (–) Output on the front
panel of the arbitrary waveform generator.
Connect an SMA cable from the Channel 1 Analog (+) Output on the
front panel of the arbitrary waveform generator to the CH 1 Input of
the sampling oscilloscope using the 12 dB SMA attenuator.
Connect an SMA cable from the Channel 1 Marker 1 (+) Output on
the front panel of the arbitrary waveform generator to the TRIGGER
DIRECT INPUT of the sampling oscilloscope using the 12 dB SMA
attenuator. (See Figure 3-4.)
b.
Set the sampling oscilloscope as follows:
Vertical External Attenuation: 12 dB
Vertical Scale: 200 mV/div
Horizontal Scale: 50 ps/div
Horizontal Timebase Mode: Lock to INT. 10 MHz
Trigger: External Direct Input, Level 50%
Acquisition: Average 64, Stop After Condition: Average Complete
Measurement: R1 to C1 Delay (R1/C1 Reference Level: Absolute
0.0 V)
Figure 3-4: Inter-channel skew calibration initial hookup
3–8
AWG7000B and AWG7000C Series Service Manual
Summary of Contents for AWG7 1B Series
Page 2: ......
Page 6: ......
Page 16: ...Environmental Considerations x AWG7000B and AWG7000C Series Service Manual...
Page 18: ...Preface xii AWG7000B and AWG7000C Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 22: ...Operating Information 1 2 AWG7000B and AWG7000C Series Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 33: ...Adjustment Procedures...
Page 34: ......
Page 52: ...Adjustment Procedures 3 18 AWG7000B and AWG7000C Series Service Manual...
Page 53: ...Maintenance...
Page 54: ......
Page 130: ...Troubleshooting 4 76 AWG7000B and AWG7000C Series Service Manual...
Page 131: ...Replaceable Parts...
Page 132: ......
Page 136: ...Replaceable Parts Figure 5 1 Cosmetics 5 4 AWG7000B and AWG7000C Series Service Manual...
Page 166: ...Replaceable Parts Figure 5 18 CLK12G module 5 34 AWG7000B and AWG7000C Series Service Manual...
Page 168: ...Replaceable Parts Figure 5 19 AWG12G module 5 36 AWG7000B and AWG7000C Series Service Manual...
Page 170: ...Replaceable Parts Figure 5 20 SMA bracket 5 38 AWG7000B and AWG7000C Series Service Manual...
Page 172: ...Replaceable Parts Figure 5 21 Relay unit 5 40 AWG7000B and AWG7000C Series Service Manual...
Page 174: ...Replaceable Parts Figure 5 22 Rear connectors 5 42 AWG7000B and AWG7000C Series Service Manual...