A6304XL Instructions
11
Maintenance
This section explains how to clean the A6304XL current probe and, if necessary,
disassemble a probe for maintenance or repair. Also included are instructions for
preparing a probe for shipment.
The A6304XL current probe should require only routine cleaning. Performance
verification and probe adjustment procedures may be found in the AM 503B &
AM 5030 Instruction Manual.
WARNING.
Probe disassembly should only be performed by qualified service
personnel.
Cleaning
To clean the probe body, use a soft cloth dampened in a solution of mild
detergent and water. To clean the core, open the jaw and clean the exposed core
surfaces with a cotton swap dampened with isopropyl alcohol (isopropanol) or
ethyl alcohol (fotocol or ethanol).
Do not lubricate the mating surfaces of the jaws. Any lubricant between the core
pieces should be removed with a recommended solvent.
Do not use chemicals containing benzine, benzene, toluene, xylene, acetone, or
similar solvents.
Do not use a petroleum based lubricant on the plastic. If the plastic slide
assembly requires lubrication, use a silicone based grease sparingly.
Do not immerse the probe in liquids or use abrasive cleaners.
Disassembly Instructions
The following procedures explain how to disassemble the probe body and
replace the current transformer.
WARNING.
Probe disassembly should only be performed by qualified service
personnel.
1. Unlock the probe slide.