Specifications
Mechanical, electrical, and environmental characteristics unique to the probe are
listed in tables 1 through 3. Please refer to the amplifier documentation for
additional specifications.
Table 1: Electrical Characteristics
Bandwidth (–3 dB)
DC to
≥
2 MHz
Rise Time (10% to 90%)
v
175 ns
Maximum Continuous Current
500 Amps (DC + peak AC)
Maximum Pulsed Current
700 Amps (DC + peak AC)
(See Figure 3)
Pulse Amp
@
Second Product
0.4 A
@
s (400,000 A
@
m
s)
DC Accuracy
±
3%
Frequency Derating
See Figure 4
Maximum Bare Wire Working Voltage
600 V
RMS
, CAT II
300 V
RMS
, CAT III
Insertion Impedance
See Figure 5
Typical Aberrations
<
±