background image

5812N 

SECTION  1 

OPERA TING  INSTRUCTIONS 

nstrument  Description 

The 

5B 12N 

D

u

a

l  Time  Base  is 

dual-sweep  plug-in  for 

use  with  Tektronix  5100-Series  Oscilloscopes.  The  plug-in 

features  two  time-base  generators,  and  is  capable 

of 

pro­

ducing  two  independent  sweeps  or 

a  d

el

a

y

e

d sweep.  The 

Sweep  time  base  provides  normal  sweep  rates  from  1 

microsecond/division  to  5  seconds/division; 

10  t

i

m

e

m

a

gni

f

ier  extends  the  displayed  sweep  time/division  to  100 

na

n

ose

c

o

n

d

s

The 

Swee

provides  sweep 

rates 

from  0.2 

microsecond/division  to  0.5  second/division.  The  plug-in 

also 

a

cc

epts  external  signals  to  be  displayed in lieu  of  the 

Sweep.  An  illuminated  knob  skirt  provides  direct  readout 

of  the sweep rates  and deflection  factors. 

CONTROLS 

AND 

CONNECTORS 

This  is 

brief 

description  of  the function  or  operation  of 

the  front·panel  controls 

and 

connectors.  More  detai led 

information 

is  given  under Genera! 

Information. 

DISPLAY 

MODE 

Applies  logic  levels  to  the  oscillo­

s c o pe  s y s tem  to  select  Chop 

(button  pushed 

in) 

or  Alternate 

(button  out)  time-shared  switching 

between 

vertical  plug-ins  and 

ampli­

f

i

e

r  c

h

a

n

n

els. 

A: 

Selects 

Time  Base  as  the 

oscilloscope  time  base.  B 

T

i

me 

Base 

is  locked 

out. 

B:  Selects  B  Time  Base  as  the 

oscilloscope  time  base. 

Time 

Base is  locked 

out. 

DUAL  SWEEP 

(both  the 

and  B 

buttons pushed in):  Selects  time­

shared  switching  between 

the 

and 

B time bases. 

If 

both  vertical 

plug-ins  are  active, 

the  A 

Time 

Base  is  slaved  to  the  left  vertical 

plug-in  and 

the  B 

Time  Base  is 

slaved  to  the  right  vertical 

plug­

in. 

A  INTEN-B  DLY'D 

(this button is 

dependent  upon  the 

and 

buttons): 

Button  in  permits 

dis­

p

la

y  of  the 

Time  Base,  with 

1-1 

the  B  Time  Base 

operating  con­

currently  and  appearing  as 

an 

i nt e n sifi e d  

segment  (the  A 

button  must  be  pressed  in),  or 

it 

permits  display  of  the 

8  Ti

m

Base,  starting  after  a  delay  time 

established  by  the  setting  of  the 

A  SECONDS/DIV 

switch  and 

the 

DELAY  TIME  MULT 

dial 

(

t

h

e  B 

button 

must  be 

pr

e

ss

e

in),  or  display  is  presented  with 

the 

sweep  being  intensified 

and  the 

sweep  being  delayed 

(both  the 

a

nd 

buttons  must 

be  pressed 

i

n). 

TRIG  AFTER  DELAY:  P

e

rmits 

the  B  Time  Base  to 

be 

triggered 

after  a  delay 

time  established  by 

the  setting 

of 

the 

A  SECONDS/ 

DIV 

switch  and  the 

DELAY 

TIME  MUL T  dial. 

AUTO  TR IG:  If 

triggering  signal  is 

absent  or  occurs  at 

rate 

less 

than  15  hertz,  the  sweep  genera­
t ors  revert  to  a  free-running 

mode 

(bright  baseline).  Button 

ou t  selects  Normal  Triggered 

Mode. 

Di

s

p

l

a

yed  time  base 

depends  on 

DISPLAY  MODE 

selected. 

A  SINGL 

SWP: 

Button  in  selects 

the  Single  Sweep  Mode,  allowing 
t

h

Time  Base  to  be  triggered 

only  once  until  manually  reset. 

Button  out  permits  repetitive 

sweep. 

A  RESET: 

Resets  sweep  circuits  to 

accept 

next  t

rigg

e

when  in  the 

Single  Sweep  Mode.  Terminates 

sweep 

if  in 

process  in  any  sweep 

mode. 

READY  INDI CATOR  I

n

dicat

e

when 

t

h

s

w

ee

circuit 

is  triggerable  in  Single Sweep  Mode. 

DELAYTIMEMULT 

Provides  v

a

r

i

a

bl

B·sweep 

de

l

a

between  0.20  and  10.20  times  the 

delay 

time  indicated  by  the 

Summary of Contents for 5B12N

Page 1: ...INSTRUCTION MANUAL Serial Number _ __ _ _ Tektronix Inc P O Box 500 Beaverton Oregon 97005 Phone 644 0161 Cables Tektronix 070 1141 00 1071...

Page 2: ...e privileges reserved Copyright 1971 by Tektronix Inc Beaverton Oregon Printed in the United States of America All rights reserved Contents of this publication may not be reproduced in any form withou...

Page 3: ...5B12N A POSITION B SOURCE LEFT DISPLAY TRIGGERING DELAY TIME MUll INTEN BAl A INTEN B OLY O 5B12N DUAL TIME BASE...

Page 4: ...Verso Filler Page z z...

Page 5: ...ins are active the A Time Base is slaved to the left vertical plug in and the B Time Base is slaved to the right vertical plug in A INTEN B DLY D this button is dependent upon the A and B buttons But...

Page 6: ...l plug in as the trigger signal source for the A Time Base COMPOSITE both the LEFT and RIGHT buttons pushed in Selects the signal or signals being displayed as the trigger signal source for the A Time...

Page 7: ...e displayed signal to produce a stable display An external triggering signal can be used to provide a triggered display when the internal signal is too 1 3 low in amplitude for correct triggering or c...

Page 8: ...ed on the screen The READY light goes out im mediately upon receipt of the sweep initiating trigger After the single sweep is terminated the sweep generator is locked out until again reset To prepare...

Page 9: ...hould be operated at a faster rate than the A sweep in the delayed sweep mode to avoid illogical displays The delayed sweep mode permits the differential delay time between two displayed events to be...

Page 10: ...delay time Fig 1 1 Comparison of the delayed sweep modes In each display the B sweep is delayed a selected amount of time by the A sweep APPLICATIONS General The following information describes proce...

Page 11: ...Subtract the first dial setting from the second and multiply the results by the A SECONDS DIV switch setting This is the time interval between the pulses Example Assume the first dial setting is 1 3 1...

Page 12: ...er at high apparent magnifica tion ranges The B TRIG AFTER D L Y mode provides a more stable display because the delayed display is triggered at the same point each time 1 Establish the display as giv...

Page 13: ...l F ig 4 Measuring pulse jitter 1 9 see Electrical Characteristics Measure the amou nt of hori zontal movement then multiply th is d i stance by the B SECO N D S D I V switch setting to obtain pulse...

Page 14: ...ggering A Sweep O n l y DC D I R ECT CO U P L E D At least 200 m V DC to 2 M H z A C CA PAC IT I V E CO UPLED At l east 200 m V 5 0 H z t o 2 M H z I N PUT R A N D C 1 MQ w i th i n 2 paral l e l ed b...

Reviews: