OPERATING
INSTRUCTIONS
FIRST-TIME
OPERATION
Plug
the unit into a 540-
or 530-Series Oscil
loscope
and turn the power on. Allow the instru
ment
to reach operating temperature, about 2 to
3 minutes
and free-run the sweep at 1 millisec-
ond/cm.
Turn
the MODE
switch to A ONLY
and
the A-channel POLARITY and AC-DC
switches to NORMAL and DC respectively.
Position
the trace to about +2 cm with the A-
channel
VERTICAL POSITION
control.
Turn
the MODE
switch to B ONLY and the
B-channel
POLARITY
and AC-DC switches
to
NORMAL and DC respectively. Position
the trace to about —2 cm with the B-channel
VERTICAL
POSITION
control.
Now
turn
the MODE switch to CHOPPED.
Two traces
will
appear on the crt screen. Notice
that the A-channel VERTICAL POSITION
control moves the upper trace and the B-channel
VERTICAL
POSITION control moves
the
lower
trace.
Increase the sweep speed to 100
microsecond/cm and notice
that each trace is
composed
of
many short-duration elements. The
two
channels
are being switched at approxi
mately 100 kc so
that each channel
conducts for
about
5
µsec and then is cut off while the other
channel
conducts for an equal time.
Now turn
the MODE
switch to the ALTER
NATE position. There are still two traces on
the crt screen but
the traces are no longer chop
ped
into small bits. For each sweep cycle one
channel
is conducting and
the other is cut off.
The
channels
are
switched at the end of each
sweep
cycle.
GENERAL
OPERATION
Either of
the two identical amplified channels
can
be
used independently by turning the
MODE Switch to A ONLY or B ONLY and
connecting the signal to be observed to the ap
propriate
input. The following
remarks
apply
equally
well to either amplifier channel.
Use
of
Probe
The
Type P410 probe, furnished with the 540-
Series
Oscilloscopes,
is designed
to preserve the
transient
response of this unit. This
probe in
troduces
no ringing
but
causes an additional
fre
quency-response loss of less than 1 db at 24 me.
The
Type P410
probe has a 10-to-l attenuation
ratio.
The
Type P510A probe is not suitable for use
with
the Type 53/54C Plug-In Unit and 540-
Series Oscilloscope
combination to observe fast
rising
pulses. This
probe tends to ring at about
50
me and
the wide passband of
the 53/54C —
540-Series
combination will display any ringing
that may occur.
Be
sure to check the adjustment of the probe
when
you first
connect it
to
a plug-in unit or
oscilloscope.
The probe
compensation
is
a func
tion of the shunt input capacitance of the par
ticular plug-in
unit or oscilloscope that you use
the probe with. If the compensation is incorrect
the frequency response will be affected. Touch
the probe tip to
the calibrator output connector
and display several cycles of the calibrator wave
form.
If
the top and
bottom of the displayed
square
wave
are not flat, adjust
the trimmer ca
pacitor
located either inside
the probe body or
inside
the box at the other end of the cable to
achieve correct
square-wave
response.
Input
Coupling
It is
sometimes undesirable
to display
the de
level
of
the waveform being observed. Placing
the AC-DC switch in the AC position inserts a
capacitor
in series with
the input so the de com
ponent
of
the waveform is blocked
and only the
ac
component is displayed. The
low-frequency
response
is about 2 cps when ac coupling is used.
Output
Polarity
It
will be
desirable
to invert the displayed
waveform
at
times,
particularly when using
the
dual-trace feature
of the 53/54C. The POLAR
ITY
switch has two positions. In the NORMAL
position
the displayed waveform will have
the
same
polarity as
the input signal. In
the IN
VERTED
position
the
displayed waveform will
be
turned upside down; that is,
a positive-going
pulse will
be displayed as
a negative-going pulse.
DC Balance
Adjustment
After the plug-in unit has been in use for a
period
of
time
you will
notice
that the trace will
change
position as the VARIABLE control is
rotated.
This
is
caused by
tube ageing and
the
resultant shift in
operating
potentials. To cor
rect
this
condition
rotate the VARIABLE con
trol
back and forth
and adjust the DC BAL con
trol
until
the trace
position is no longer affected
by
rotation of
the VARIABLE control.
Gain
Adjustment
Ageing
of the tubes will also affect
the gain
of
the plug-in unit. Display
a calibrator wave
form of 0.2 volt peak to peak with the
VOLTS/CM
switch in the .05 position. Adjust
the GAIN ADJ control until the displayed
waveform
is 4 graticule divisions in amplitude.
Make
sure
the
VARIABLE control is turned
full
right to
the CALIBRATED position before
making
this adjustment.
53/54C
—
OPERATING
INSTRUCTIONS
—
PAGE
1
Summary of Contents for 53C
Page 2: ...B http manoman sqhill com...
Page 17: ...4963X1 _L T C 53 54C ATTENUATION NETWORKS 4 19 57 K F...
Page 18: ......