Performance Check/Calibration—Type 324
Square-w ave
generator
High-frequency
Constant-am plitude
sine-w ave generator
Standard
am plitude
C alib rato r
Tim e-M ark
generator
Low-frequency
Constant-Am plitude
sine-w ave g enerator
C alib ratio n
shield
v>M **•(<*■
W f t f f i f l
nri 11 m
Test
oscilloscope
fc*
P 6049
probe
42-inch
50-ohm
BNC C ab le
BNC T
Connector
50-ohm
BNC
term ination
Low -Capacitance
screw driver
1 8-inch
50-ohm
BNC cable
Six-inch
b a n a n a plug
to BNC patch
Cord
1 0 X BNC
attenuator
GR to BNC
ad ap ter
4 7 pF
input RC
norm alizer
mm
'4m
'
■
.
:
BNC post
jack
'/■t
.•;> •'.<
Three-inch
screw driver
Fig. 5-7. Test equipment required for steps 21 through 46.
Partial Procedure
If beginning a partial procedure with this step, set
the controls as given under Preliminary Control Settings.
21. Check Trace Shift Due to Input Current
a. Test equipment required for steps 21 through 46
is shown in Fig. 5-7.
b. Change the following control settings:
INPUT
GND
X5 VERT GAIN
Pulled out
c.
Install the calibration shield (067-0622-00) on the
Type 324.
d.
Position the trace to the center horizontal line
with the vertical POSITION control.
e.
CHECK—Set the INPUT switch to DC. Trace shift
should be negligible (0.085 division maximum).
f. Remove the calibration shield.
22. Check/Adjust High-Frequency
O
Compensation
a. Change the following control settings:
X5 VERT GAIN
Pushed in
TIM E /D IV
1 ps
X5 HORIZ MAG
Pulled out
b.
Connect the square-wave generator (Type 106)
fast-rise + output connector to the VERT INPUT con
nector through the GR to BNC adapter, 42-inch 50-ohm
BNC cable, 10X BNC attenuator and the 50-ohm BNC
termination.
c.
Set the square-wave generator for a four-division
display at 100 kilohertz.
d.
Move the leading edge of the square wave onto
the viewing area with the horizontal POSITION control.
5-17