![Tektronix 067-0521-00 Instruction Manual Download Page 9](http://html.mh-extra.com/html/tektronix/067-0521-00/067-0521-00_instruction-manual_1078800009.webp)
Operating Instructions— 0 6 7 -0 5 2 1 -0 0
5. Carefully focus and check the display. If the waveshape
does not show good transient response because of over
shoot, ro llo ff or bumpiness, proceed with the adjustments
described in the oscilloscope instruction manual.
Other Checks
The 067-0521-00 can be used as a limited bandpass plug-in
unit. It is useful for inserting calibrated time marks into
the vertical system when adjusting the oscilloscope geometry,
trigger, and time-base circuits for proper operation. For
making a sine-wave bandpass check of an oscilloscope pre
am plifier system, use a 1 -series or letter-series plug-in unit
rather than the plug-in test unit.
When using the 067-0521-00 to couple an input signal to
the oscilloscope vertical amplifier, the deflection factor of
the unit is about 0.5 volt/cm with the TEST FUNCTION
switch set to LOW LOAD or HIGH LOAD and the VARI
ABLE control turned fully clockwise. LOW LOAD is the
normal position of the TEST FUNCTION switch for dis
playing signals applied through the EXT INPUT connector.
Resistance Measurements
Blank columns are provided in Table 4-1 for recording
resistance readings and the meter used. Resistance measure
ments recorded when the 067-0521-00 operates correctly may
be useful if trouble should occur.
®
2 -3
Summary of Contents for 067-0521-00
Page 10: ...NOTES ...
Page 24: ...NOTES ...
Page 46: ...CIRCUIT DIAGRAM x x r s 3 O z z l ...
Page 47: ...J 067 0521 00 CALIBRATION FIXTURE ...
Page 49: ......
Page 51: ...I ...
Page 52: ...TYPE 0 6 7 0 5 2 1 0 0 CHANGE R5 f PARTS LIST CORRECTION 321 0152 00 37 Q 1 8 w M12 536 967 ...
Page 53: ......