Socket Cables and

Solder Tip Kits

ZZZ

Instructions

Overview

These instructions apply to the following kits:

Kit part
number

Description

Qty

020-2954-00 Socket Cable

1

020-2960-00 Socket Cable XL &

DSP Filter Files CD

1

020-2958-00 TriMode High Temp Solder Tip

10

020-2955-00 TriMode Micro-Coax Solder Tip

10

020-2959-00 Damped Wire Pair

25

Order the resistor kit below to replace the
resistors on your solder tips.

020-2937-00

TriMode Solder Tips Replacement
Resistor Kit (Includes 50 each of
the following components)
100

leaded resistor

75

surface-mount resistor, 0402

Non-conductive tube

Description

The Socket Cables and Solder Tips allow you
to create customized, hands-free connections
from your circuit to Tektronix P7500 Series
TriMode probes. Integral resistors on the tips
include small-diameter, pre-trimmed leads for
you to solder to your circuit board vias or other

fi

ne-pitch features. Sockets at the ends of the

cables allow you to quickly change between the
soldered-down tips at multiple test points.

The TriMode High Temp Solder Tip and
TriMode Micro-Coax Solder Tip support full
TriMode capability (differential, single-ended,
and common-mode con

fi

gurations). The

Damped Wire Pair are designed for differential
measurements only.

The Socket Cable XL is 5 feet long and is
intended for use only with the TriMode High
Temp Solder Tip. Always use the DSP

fi

lter

fi

les

when using the XL cable.

Using the Solder Tips

General recommendations:

Use the Damped Wires only in pairs, in the
A-B (differential) mode.

The solder tips cannot support the weight of
the attached probe, so you must secure the
probe to the circuit board to relieve strain on
the tips and the solder connections.

Use a

fl

ux pen to add

fl

ux to the solder tips

and to your test points. The

fl

ux makes

soldering easier.

Use only ESD-approved soldering irons and
no-clean

fl

ux solder when soldering.

Procedure

1.

Choose a location where the solder tip leads
can reach your test points, while keeping
the tip leads as short as possible for the best
signal quality. Note signal polarities and
grounds where applicable.

2.

Apply solder to the test points on your
circuit.

3.

Solder the tip leads to the test points on your
circuit and clip off any excess.

4.

Connect the pins of the tip to the sockets
in the cable. Note signal polarities and
grounds where applicable.

5.

Align the cable with the red band to the A
input of the probe. Push the cable end into
the probe head until it seats in the head.

6.

For a secure mechanical connection, use
double-sided tape or hot glue to secure the
tip to your circuit.

Signal Quality with the Socket Cable XL

Due to the length of the Socket Cable XL, the
input signal to the oscilloscope requires some
DSP

fi

ltering. You must load

fi

lter

fi

les onto your

oscilloscope from the CD that is included in your
XL cable kit. Always use the

fi

lter

fi

les when

you use the Socket Cable XL.