013-0278-00 Video Display Clamp
7
Performance Verification
This
Performance Verification
subsection confirms the performance of the
013-0278-00 Video Display Clamp to its warranted specifications.
General Instructions
These
General Instructions
describe the conventions used in the test procedures
and provides a list of the required test equipment.
The procedures in this section provide the following information:
H
Title of test
H
Equipment required (if applicable)
H
Procedure
The procedures check all the characteristics designated as checked in the
Specifications
subsection. (The checked characteristics appear in
boldface
type.)
These procedures use external, traceable signal sources to directly check
warranted characteristics. Table 7 lists the required test equipment.
Table 7: Test Equipment
Item Number and Description
Minimum Requirements
Example
Purpose
1.
Oscilloscope
TEKPROBE
TM
Level 2 interface,
100 MHz bandwidth
Tektronix TDS460
Signal display
2.
Cable, Precision Coaxial
50
Ω
, 36 in, male to male BNC connec-
tors
Tektronix part number
012-0482-00
Signal interconnection
3.
Termination, 50
Ω
Impedance 50
Ω
; connectors: female
BNC input, male BNC output
Tektronix part number
011-0049-01
Signal interconnection
4.
Attenuator, 2.5X
Ratio: 2.5X; impedance 50
Ω
; connec-
tors: female BNC input, male BNC output
Tektronix part number
011-0076-02
Signal attenuation
5.
Coupler, Dual-input
(two required)
Female BNC to dual male BNC
Tektronix part number
067-0525-02
Signal interconnection
6.
Generator, Leveled Sine
Wave
30 MHz; Variable amplitude from 5 mV to
4 V
p-p
into 50
Ω.
TEKTRONIX SG503 Lev-
eled Sine Wave Genera-
tor
1
Bandwidth check
Conventions
Equipment Required