9 of 12
© 2010 D 132 - 04/10
The Pump Sequencer 132 has a built-in test routine which is used to test
the main control functions. The 132 continually monitors the outdoor sensor
off not testing (if installed) and displays an error message whenever a fault
is found. See red testing the following page for a list of the 132’s error
messages. When the
Test
red testing paused button is pressed, the test
light is turned on. The individual outputs and Test relays are tested in the
following test sequence.
TEST SEQUENCE ------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------
Each step in the test sequence lasts 10 seconds.
During the test routine, the test sequence is paused by pressing the
Test
button. If the
Test
button is not pressed again for 5
minutes while the test sequence is paused, the control exits the entire test routine. If the test sequence is paused, the
Test
button
can be pressed again to advance to the next step. This can also be used to rapidly advance through the test sequence. To reach
the desired step, repeatedly press and release the
Test
button until the appropriate LED is turned on.
Step 1
-
Pump 1
contact is turned on for 10 seconds. After 10 seconds, the
Pump 1
contact is shut off.
Note:
Only if there is a pump demand can the control be paused in step 1.
Step 2
-
Pump 2
contact is turned on for 10 seconds. After 10 seconds, the
Pump 2
contact is shut off.
Note:
Only if there is a pump demand can the control be paused in step 2.
Step 3
-
Alert A
contact is turned on for 10 seconds. After 10 seconds, the
Alert A
contact is shut off.
Step 4
-
Alert B
contact is turned on for 10 seconds. After 10 seconds, the
Alert B
contact is shut off.
ERROR HANDLING ----------------------------------------------------------------------------------------------------------------------------------
----------------------------------------------------------------------------------------------------------------------------------
If an internal control fault occurs, the 132 displays an error message and turns on the
Alert A
contact. In this case the control
continues operation. To clear the error message from the control, press the
Clear Alert
button. If the error message remains, the
control must be returned for repair.
If an outdoor sensor fault occurs and the WWSD dial is not set to Off, the 132 displays an error message and turns on the
Alert A
contact. In this case the control operates the pumps based on demands. Locate and repair the problem as described in the Data
Brochure D 070. To clear the error message from the control after the sensor has been repaired, press the
Clear Alert
button.
If a flow proof signal is present when both pump contacts have been off for more than 4 minutes, the 132 displays an error message
and turns on an alert contact. In this case the control will stop operation. To clear the error message from the control after the fault
is corrected, press the
Clear Alert
button.
If a pump fails, the appropriate error message and alert signal is displayed. To clear the error message from the control after the
fault is corrected, press the
Clear Alert
button.
If the 2-stage demand short cycles, the 132 displays an error message and turns on the
Alert A
contact. In this case the control
continues operation. To clear the error message from the control after the fault is corrected, press the
Clear Alert
button.
Testing the Control
not testing
testing
testing paused
of f
red
red