Section IV - Operating Instructions
PS-CAL in Batch Mode
PS-Cal is able to calibrate power sensors in batch mode. Though PS-Cal can only calibrate a single sensor
at a time, Batch Mode allows the operator to configure that station to test one power sensor, and then use
that configuration to calibration the next sensor in the batch. This process saves time, because many
power sensor calibration configuration steps are time consuming.
Figure 4.11 Batch Mode
To use PS-Cal in Batch Mode:
1.
Click on UUT in the menu bar
2.
Select New.
This adds an Asset tab.
3.
Label the tab by selecting it, and then selecting the UUT tab, and entering the asset number in the
Asset Number field.
4.
The new asset tab is now labeled.
You can now select the newly added asset by clicking its tab along the left. Configuration information and
data collected on one power sensor is shared with all the power sensors that are open.