
Service Information
4
-
6
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
Troubleshooting
TEGAM’s digital handheld thermometers are designed and built to provide years of
uninterrupted use. In the event the instrument malfunctions or does not perform as
expected, helpful troubleshooting tips are provided below.
Figure 14
below lists some of the
more common issues and their resolutions.
S
YMPTOM
D
ESCRIPTION
R
ESOLUTION
Unexpected
reading
on Line 2
of Display
Statistics View Mode
is
active
Press
to cycle through statistics
views until active measurement is
displayed
(see
Section 3.4 View
Unexpected or
Erroneous
Measurement
Probe offset is active
Set probe offset to correct value for
connected temperature probe (see
Section 3.10, Probe Offset)
Temperature probe has
not stabilized
Observe display
trend indicators and
wait for stable measurement (see
Section 3.8 Trend Indicators
Instrument is set to the
wrong thermocouple type
for the attached probe
Set the thermocouple type as
appropriate for the attached probe
(see
Section 3.3, Setup Menu
When sourcing from a
thermocouple simulator,
Open Wire
Detection is
enabled
.
See
Section 3.11, Open Wire
to disable.
Unresponsive
Hold Mode is active
Press
, and verify that the
HOLD
indicator is not active (see
Section
Static discharge through
connected probes
Press
to cycle instrument power
Shuts down
unexpectedly or
will not power on
Batteries are low or
depleted
Replace batteries (see
Section 2.7,