Operating Instructions
3-8
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
a.
Alternatively, to disregard the new offset value and exit the Setup Menu,
press
.
9.
OFFSET
is displayed at the top
-
left of the LCD display.
To set the probe offset when using a calibrated temperature probe with a known offset:
1.
to enter the Setup Menu;
2.
Press
three (3) times to cycle to the Channel 1 Offset parameter;
3.
Observe the current offset value displayed on the second line of the display;
4.
Press
to set the offset in 0.1 ° increments until the displayed offset value
equals the calibrated probe offset value
;
5.
Press
to save the offset value and proceed to Channel 2 offset (
922B
only), or
press
to save the offset value and exit the Setup Menu.
a.
Alternatively, to disregard the new offset value and exit the Setup Menu,
press
.
6.
OFFSET
is displayed at the top
-
left of the LCD display.
Open Wire Detection On/Off
Open Wire
Detection
allows the unit to detect if a thermocouple probe is connected
to the
thermometer.
This feature is not compatible with
some
thermocouple
calibrators
and
can
result in measurement instability.
Turning Off
Open Wire
Detection in these situations can significantly improve reading
stability. Once off, Open Wire
Detection will remain off
until changed by following the below
steps, or the instrument is powered off.
If no thermocouple probe is connected and Open Wire
Detection is disabled,
the unit
will not indicate OPEn
and may display erratic readings.
To change the Open Wire
Detection setting:
1.
to enter the Setup Menu;
2.
Press
four (4) times for
921B
, five (5) times for
922B
, to cycle to the Open
Wire
Detection Off
/
On
parameter;
3.
“O
W
D
OFF
” is
flashing near the bottom of
the LCD
display, and the current Open
Wire
Detection status
is displayed on Line 2.
4.
Press
to change the Open Wire
Detection setting
as shown on Line 2 of the
display
;
a.
ON indicates that
Open Wire
Detection
is enabled;
b.
OFF indicates that
Open Wire
Detection
is disabled;
5.
Press
or
to save the
Open Wire
Detection setting
and exit the Setup Menu.