Operating Instructions
3-10
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
The CEIL function provides 1° resolution and is used when approaching the target
temperature moving towards zero in either direction, up or down.
The Trunc function provides 1° resolution and is used when approaching the target
temperature moving away from zero in either direction, up or down.
See detailed descriptions below.
rndG
1.
AutO: 0.1° Resolution for readings <1000° and 1°>999.9°. AutO is the default
setting.
2.
CEIL: 1° Resolution with ceiling rounding is used for a decreasing positive
temperature to assure the lower temperature is fully reached or an increasing negative
temperature to assure a higher temperature is fully reached.
For Example: A decreasing (cooling) temperature 20.4° actual temperature will display 21°
until 20.0° is reached. An increasing (thawing) temperature -20.6° will display -21° until -
20.0° is reached.
3.
nOrL: 1° Resolution with normal rounding, 20.4° displays 20°, 20.5° displays 21°
4.
trnC: 1° Resolution with truncate rounding is used to assure that an increasing
positive temperature is fully reached or a decreasing negative temperature to assure the
lower temperature is fully reached.
For Example: An increasing temperature (heating) 20.6° will display 20° until 21.0° is
reached.
A decreasing temperature (cooling) -20.6° will display -20° until -21.0° is reached.
This is for data on the instrument display only. Data transferred via the TEGAM
Thermometer link to the TEGAM Cloud or via Bluetooth from the instrument will not have
any rounding applied.