Section II – Preparation for Use and Installation
Unpacking & Inspection
Each Coaxial RF Power Standards is put through a series of electrical and mechanical inspections before
shipment to the customer. Upon receipt of your instrument unpack all of the items from the shipping
carton and inspect for any damage that may have occurred during transit. Report any damaged items to
the shipping agent. Retain and use the original packing material for reshipment if necessary.
Upon Receipt, inspect the carton for the following items:
Table 2.1 Packing List
Item
Part Number
Model 2505A or 1505A Coaxial RF Power
Standards
2505A or 1505A
Technical (Operation and Maintenance) Manual
1505A/2505A-900
Calibration Certificate with Data
N/A
Mounting
The Model 2505A is shipped with four plastic feet mounted to the bottom cover. When any of these
models are placed on a bench or table, the feet support the instrument. For rack mounting of Model
2505A use rack mount kit, (Single) P/N 1830-910 – (Dual) P/N 1830-911.
The Model 1505A is shipped without any type of rack mounting equipment the new design allows for use
as a terminating sensor and its low weight of 544.3 g (1.2 lb) reduces any stress on connectors.
Only use with Proper Equipment:
Connecting this device to any other type of circuit or may result in permanent damage to the components.
Additionally, the heater elements used for the internal heater are designed to be used only with the heater
control circuits found in TEGAM Models 1830A, 1820B, 1806A, 1806, and 1805B. Connecting these heater
elements to any other circuit or device may result in damage. Contact TEGAM for any questions regarding
instruments that are compatible with these RF Power Standards.
Use in Proper Environment
Normal calibration laboratory practice dictates that the environment should be closely controlled. This will
minimize errors introduced by temperature and humidity changes. A nominal temperature of +23°C
(+73.4°F) provides a good working condition. A tolerance of ±1°C gives an ideal temperature spread.
Controlled temperatures also stabilize the aging process of the standards.
CAUTION: The RF Power Standards have a specified operating temperature range of +15° to +30° C
(+59° to +86° F). Operating beyond these limits can affect the accuracy of the instruments and damage
internal circuitry.
CAUTION: When an RF Power Standard is to be stored for extended periods, pack the instrument into a
container. Place container in a clean, dry, temperature-controlled location. If instrument is to be stored in
excess of 90 days, place desiccant with items before sealing container. The safe environmental limits for
storage are -55° to +75° C (-67° to +167° F) at less than 95% non-condensing relative humidity.