Tecnosens NCL C10 Assembly Instructions Manual Download Page 28

5. Troubleshooting and Spare Parts  

24 

 

 

Problem detected 

Possible cause 

Solution 

Deformed or 
broken membrane 

Extraction or insertion of 
the probe inside the 
probe holder without 
having opened the bleed 
valve 

Membrane cap replacement. See spare parts 

Incorrect assembly or 
disassembly of the cap 

Membrane cap replacement. See spare parts 

Operating pressure 
greater than 1 bar 

Install  a  pressure  regulator  upstream  of  the 
probe holder or move the sampling point to a 
system  position  where  the  pressures  are 
compatible  with  the  operating  conditions  of 
the probe. 

Presence of solid 
abrasive particles in the 
water flow 

Install a filter upstream of the probe holder to 
prevent  particles  from  coming  into  contact 
with the probe. Membrane cap replacement. 
See spare parts 

Presence of 
bubbles on the 
membrane surface 

Flow rate and pressure of 
the sample water outside 
the operating range (30 ÷ 
60 l/h and 0.3 ÷ 0.5 bar) 

If  the  pressures  are  greater  than  those 
indicated, insert a valve to adjust the pressure 
upstream  of  the  probe  holder.  Replace  the 
membrane  cap.  See  spare  parts.  If  the 
pressures are lower than those required, try to 
throttle the system with a valve downstream 
of the probe holder.  

Measurement 
deviation from the 
DPD value 

Probe installed for too 
short a time. 

Wait another time, repeat the calibration after 
a few hours. 

Damaged membrane 
(deformed, broken, dirty, 
etc.) 

Replace  the  damaged  membrane  cap  and 
restore the gel electrolyte. See spare parts. 

Presence of chemicals 
that can interfere with 
the measurement 

Evaluate the presence of interferents present 
in the measurement water and remove them. 
If necessary, contact the supplier. 

Reference electrode 
touched with the fingers 
before fitting the cap 
with the gel electrolyte. 

Return  the  probe  to  the  supplier  for 
regeneration 

Contaminated or 
depleted gel 
electrolyte 

Tubular gasket deformed 
to the point of no longer 
guaranteeing the tightest 
seal on the cap holes. 

Replace the tubular gasket. See spare parts. 

Summary of Contents for NCL C10

Page 1: ...B EN 05 2021 Free and Total Chlorine Sensors NCL and NCT model PLEASE READ CAREFULLY THE FOLLOWING INSTRUCTIONS FOR USE DO NOT THROW THEM AWAY THE SYSTEM MANAGER WILL BE RESPONSIBLE FOR ANY DAMAGE CAUSED BY INSTALLATION OR MANAGEMENT ERRORS ...

Page 2: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...

Page 3: ...TMENT 10 3 ELECTRICAL CONNECTIONS 12 3 1 ELECTRICAL INSTALLATION OF VOLTAGE OUTPUT 12 3 2 ELECTRICAL INSTALLATION OF CURRENT OUTPUT 14 4 ACTIVATION AND MAINTENANCE OF THE SENSOR 16 4 1 SENSOR COMMISSIONING 16 4 2 CALIBRATION 16 4 3 ELECTRODE UNIT AND CAP MAINTENANCE 18 4 4 PUT THE PROBE OUT OF SERVICE 20 4 5 WORKING PARAMETERS AND SENSOR CHARACTERISTICS 21 5 TROUBLESHOOTING AND SPARE PARTS 22 5 1 ...

Page 4: ......

Page 5: ...trical output for the different sensors to which this manual can be applied The range of sensors for the measurement of free and total chlorine has a structure of the type with three electrodes two of which are covered by a membrane cap and immersed in a gel electrolytic solution The range allows to measure the concentration of free total chlorine in surfactant free water These devices can be used...

Page 6: ...e 0 05 20 00 0 01 0 2000 mV NCT T10 Total Chlorine 0 05 10 00 0 01 0 2000 mV NCT T5 Total Chlorine 0 05 5 00 0 01 0 2000 mV NCT T2 Total Chlorine 0 01 2 00 0 005 0 2000 mV NCT C20 Total Chlorine 0 05 20 00 0 01 4 20 mA NCT C10 Total Chlorine 0 05 10 00 0 01 4 20 mA NCT C5 Total Chlorine 0 05 5 00 0 01 4 20 mA NCT C2 Total Chlorine 0 01 2 00 0 01 4 20 mA NCL T20 SW Free Chlorine 0 05 20 00 0 01 0 2...

Page 7: ... as shown in Figure 1 and in Figure 2 in which the standard kit consists of I sensor II upper element not present in the version with voltage output III membrane cap IV bottle of gel electrolyte 100ml V paper for polishing the gold electrode VI user manual Figure 1 version with current output standard supply ...

Page 8: ...1 Introduction 4 Figure 2 version with voltage output standard supply ...

Page 9: ... composing the sensor are shown in Figure 3 Position Description 1 Sensor body 2 Connector 3 Counter electrode 4 Reference Electrode 5 Working electrode 6 O Ring 7 Cap 8 Membrane 9 Tubular gasket Figure 3 Sensor structure and description of its components ...

Page 10: ... cap with the gel electrolyte solution contained in the bottle present in the package The gel electrolyte must reach the edge of the cap as reported in Figure 4 leaving the tubular gasket displaced from its place Insert the probe vertically into the cap as shown in Figure 5 on the next page taking care during the assembly phase not to cover the excess gel outlet holes with your fingers This is in ...

Page 11: ...he gel solution in the cap For this reason it is useful to slowly screw the cap to the probe and not to shake the bottle before use After filling or restoring the electrolyte solution the probe takes about an hour to resume normal operation RESTORING THE ELECTROLYTIC SOLUTION It is recommended to calibrate the instrument to which the probe is connected after the operation has been restored It is r...

Page 12: ...d lead to malfunction of the probe 2 3 INFORMATION ON THE ELECTROLYTE GEL KEEP THE PRODUCT OUT OF REACH OF CHILDREN ELECTROLYTE GEL DO NOT SHAKE THE BOTTLE BEFORE USE The electrolyte is sensitive to oxidation always keep the bottle of electrolyte closed after use Do not transfer the electrolyte into other containers The electrolyte must not be stored beyond the permitted terms for the expiry date ...

Page 13: ...re available on the manufacturer s website at http www aedes info prodotto asp 2 4 MOUNTING ON THE PROBE HOLDER Insert the probe into a dedicated probe holder in the appropriate housing A distance of about 2 cm must be ensured between the membrane and the bottom of the probe holder as indicated in Figure 6 on the next page in order to ensure optimal operation of the probe The phase of installing t...

Page 14: ... the probe the flow rate of fluid passing through the probe holder must be between 30 l h and 60 l h and with a pressure between 0 bar and 1 bar the recommended working range for probe is 0 3 0 5 bar WATER FLOW Do not exceed the recommended working range do not use the sensor with a lower or higher range than the indicated limits before commissioning the probe the flow rate must be measured using ...

Page 15: ...2 Installation 11 Figure 6 installation of the probe in a probe holder where the optimal operating distance is shown ...

Page 16: ...measuring instrument to use the dedicated electric cable see paragraph 5 1 page 21 to identify the code to be purchased separately The cable is made up on one side of the connector and on the other of the wires as in Figure 7 to be connected as shown in Table 2 For the electrical installation of the probe it is necessary to connect the cable to the measuring instrument and then couple the connecto...

Page 17: ...tions Terminal Color Function Value 1 Green Signal Vout 0 2000 mVdc 2 Yellow Signal GND 0 3 Brown Power supply Vin 10 30 Vdc 4 White Power supply GND 0 Single Power Supply Terminal Color Function Value 1 Green Signal Vout 0 2000 mVdc 2 Yellow Signal GND 0 3 Brown Power supply Vin 5 15 Vdc 4 White Power supply Vin 5 15 Vdc Single Power Supply ...

Page 18: ...onnector and pass the measuring cable 1 through it 3 Uncover the ends of the cable by 0 5 cm apply the cable lugs ϕmax 0 5 mm2 and connect them to the 2 conductor connection respecting the polarity indicated on the terminal 4 4 Fully insert the upper element of the sensor 3 on the probe body 5 and tighten it by turning clockwise until it stops 5 Insert the measurement cable as far as possible into...

Page 19: ...4 Activation and Maintenance of the Sensor 15 Figure 8 Electrical installation for probes with two wire current output ...

Page 20: ...red for the probe to provide a stable signal Phase Time h First Installation 1 24 Subsequent installations 1 24 Gel and or Cap replacement 1 3 4 2 CALIBRATION The probe calibration must be carried out once the probe signal is stable therefore not before the activation times indicated in Table 3 The calibration of the measuring instrument connected to the probe must be carried out and checked perio...

Page 21: ...PD 1 measurement in the measuring instrument to which the probe is connected respecting the procedure provided in the instrument manual If it is necessary to carry out the zero calibration of the instrument 1 Remove the sensor from the probe holder taking care not to disconnect it from the power supply 2 Place the sensor in a container containing clean water free of chlorine and oxidants e g natur...

Page 22: ... recommended to clean it using the special paper in the package The paper has a shiny and a matte side the matte side is the one to be used on the working electrode as shown in Figure 10 In the event that this operation does not lead to improvements please contact the assistance center Do not use the paper on the membrane or other parts MAINTENANCE INFORMATION To ensure correct operation of the pr...

Page 23: ...ivation and Maintenance of the Sensor 19 Figure 9 Working electrode before and after cleaning Figure 10 Cleaning the working electrode with abrasive paper Figure 11 Do not use abrasive paper on the membrane ...

Page 24: ...Do not touch the electrodes with your fingers 7 Leave the cap and the probe to dry in a place protected from dust Do not use air jets 8 To protect the electrodes during storage screw the cap to the probe without tightening 9 To protect the sensor keep it in its specific box INSTALLATION INSTRUCTIONS Before reactivating a probe that has been taken out of service for a long time it is recommended to...

Page 25: ...ture must not exceed 45 C If the water temperature is close to 0 C it is essential for the operation of the probe that there are no ice crystals in the flow The value measured by the sensor is not very dependent on the pH The pH range optimal for operation is between 4 pH and 10 pH The reference electrode can only be cleaned with water and left to dry on its own without the aid of cloths or paper ...

Page 26: ...trol instrument Power line and electrical connections Hydraulic line and related fittings Sensor Before starting the troubleshooting make sure that the operating conditions below are met If the measured values of the sensor deviate significantly from the measured value of the DPD method all possible errors of the photometric DPD method must first be considered If necessary repeat the DPD measureme...

Page 27: ... the regeneration Contaminated or depleted gel electrolyte Replace the gel in the cap with the one supplied if it is exhausted see spare parts Check the integrity of the reference electrode Presence of air bubbles on the membrane Replace the gel in the cap with the one supplied if it is exhausted see spare parts Check the integrity of the reference electrode Incorrect calibration of the instrument...

Page 28: ...e 30 60 l h and 0 3 0 5 bar If the pressures are greater than those indicated insert a valve to adjust the pressure upstream of the probe holder Replace the membrane cap See spare parts If the pressures are lower than those required try to throttle the system with a valve downstream of the probe holder Measurement deviation from the DPD value Probe installed for too short a time Wait another time ...

Page 29: ...rewed cap which caused water infiltration and or leakage of gel electrolyte Clean the working electrode and replace the electrolyte gel Carefully screw the cap back on and reposition the tubular gasket correctly It may be necessary to re calibrate the instrument to which the probe is connected ...

Page 30: ...ERVED Reference EC directives EMC Directive 2014 30 EU ROHS Directive 2011 65 EU International standards EN 61010 1 EN 60335 1 EN 60529 EN 61326 1 NOTICE We reserve the right to make changes at any time and without notice in colors materials specifications and designs ...

Page 31: ...7 Notes 27 7 NOTES ...

Page 32: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...

Reviews: