5. Troubleshooting and Spare Parts
22
5
T
ROUBLESHOOTING
A
ND
S
PARE
P
ARTS
5.1
S
PARE
P
ARTS
Code
Description
NAS0003A2
Cap for probes NCL
NAS0006A0
Cap for probes NCT
NSH0001A1
Gel electrolyte bottle (100ml) for NCL
NSH0002A0
Gel electrolyte bottle (100ml) for NCT
NAS0005A1
2 m cable for probes with voltage output
5.2
T
ROUBLESHOOTING
For troubleshooting, it is necessary to consider the entire measuring station.
The measuring station consists of:
•
Measurement and control instrument
•
Power line and electrical connections
•
Hydraulic line and related fittings
•
Sensor
Before starting the troubleshooting, make sure that the operating conditions
below are met.
If the measured values of the sensor deviate significantly from the measured
value of the DPD method, all possible errors of the photometric DPD method
must first be considered. If necessary, repeat the DPD measurement several
times.
Table 4: Sensor operating condition summary
Parameter
Value
Concentration of chlorine
Dependent on the full scale of the probe
pH
4.0 ÷ 10.0, Constant
Temperature
0 °C ÷ 45 °C, without temperature
changes
Conductivity
0.1 ÷ 63 mS / cm
Flow rate
30 ÷ 60 l/h
Summary of Contents for NCL C10
Page 2: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...
Page 4: ......
Page 8: ...1 Introduction 4 Figure 2 version with voltage output standard supply ...
Page 31: ...7 Notes 27 7 NOTES ...
Page 32: ...VIA FRANCESCO ANTOLISEI 25 00173 ROME ITALY 39 06 4576 3210 www tsens biz info tsens eu ...